If the force transducer in the stage is changed for any reason (such as swapping between the 20N and 200N transducer) a distance calibration must
be performed. To begin this, tap on the 'Length' button, which will pop up a confirmation dialogue:
Make sure the jaws are free of obstructions and tap the OK button to continue, which will return you to the main control screen while the calibration
takes place. This process is automatic, the tensile displays will change to 'Cal' versions while this is happening, as below:
While these readings show 'DCal' and 'FCal' do not operate the stage, the readings will revert to normal once the calibration is complete and the
system is ready for use.
Important note:
The Force calibration button is only for use by Linkam Scientific and requires a password to access.
Do not attempt to use this
control.
DSC stages
This section applies to the DSC600, DSCX600 and DSC450 stages only.
The only additional information displayed for a DSC600 or DSCX600 stage is the DSC count (1). There are no additional controls. Please see your
DSC600 stage manual for an explanation of DSC counts.