39
To set up the alarm volume:
Press button
6
until the desired volume is reached.
NOTE: If the alarm is set to minimum volume the mute mode is on.
Inner Zone alarm (indicator 9)
Alarm starts when patient moves from the limited area.
Outer Zone alarm (indicator 8)
Alarm starts when patient leaves bed.
NOTE Inner zone alarm is the default mode when the Bed Exit Alarm is activated.
7) Bed Overload
If the bed load is over 254,5kg:
Overloading is signalized by long acoustic signal.
The „
Hi
“ icon is displayed on the display.
NOTE: If the bed is overloaded then it is impossible to position or manipulate with the bed until the
overloading is removed.
NOTE: The bed overloading has always higher priority than Hold Mode and Taring functions.
8) Bed Underload
In case the bed is underloaded (factory zero
– 5kg):
The display shows icon „
Lo
“.
9) Weighing in tilt
The bed can weight in tilt. The guarantee of accuracy is secured by spirit level which is located at the head/foot end
of the bed. If the bubble is in the highlighted circle then the weighting is accurate.
10) Zeroing Scales
The zeroing can be done only in the range ±5kg from the factory zero. The zeroing is used to reset weight on the
display and also to set up user zero, which sets the maximum weight range of the weighting system.
The zeroing must be done on the empty, unloaded bed without mattress and accessories. The zeroing is done
after installation, weight verification or servicing.
To zero scales:
Remove all accessories and mattress from the bed. Position the bed about 20 cm above the lowest
position and the mattress platform to the horizontal position. Ensure that nothing touches the bed except
you.
Press button
3
(Zero/T) for 0,5s until weight value starts to flash.
Press button
3
to confirm zeroing.
„0“ is shown at the display and acoustic signal confirms zeroing.
To cancel zeroing:
Press button
4
while zeroing.