![Lindberg/Blue Mini-Mite TF550 A Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/lindberg-blue/mini-mite-tf550-a/mini-mite-tf550-a_installation-and-operation-manual_1919398013.webp)
Mini-Mite™ Tube Furnace
UP150 Controller Operation
10
In the table below, ‘*' denotes values typically set by the user according to the needs of a program.
Parameter
Value
Meaning
Parameter
Value
Meaning
PrG
0
Enters program menu
SP4
*
Segment 4 Setpoint
SSP
25
Start Setpoint
Tm4
oFF
Time length for Segment 4
StC
*
Start Code
EV1
0
Event 1
SP1
*
Segment 1 Setpoint
AL1
9
Alarm 1
tm1
*
Time length for Segment 1
1
1
Hysteresis for alarm 1
SP2
*
Segment 2 Setpoint
0
0
Event 2
tm2
*
Time length for Segment 2
AL2
oFF
Alarm 2
SP3
*
Segment 3 Setpoint
JC
1
Junction code (1=dwell; O=stop;
2=repeat)
tm3
*
Time length for Segment 3
wtz
oFF
Wait Zone
7.3.3 Running a Program
To run a program such as the one outlined above, press and hold the 'DOWN/RUN' button making the ‘RUN’ indicator illuminate.
At the end of this program the ‘HLD’ (hold) indicator is illuminated to indicate this program is in the indefinite dwell at the last target
temperature. This hold indicator is caused by the "JC" selection of "1", while the "JC" selections of "0" or "2" will not illuminate the
‘HLD’ (hold) indicator.
7.3.4 Ending a Program
To end a program while in the 'RUN' or 'HLD" (hold) mode, press and hold the 'UP/RESET' button to turn off the current program
and extinguish the 'RUN' or 'HLD' indicator.
Turning off the unit’s power will also stop the program. When power is restored, the controller is in the Reset or standby mode with
no power to the heaters.
Note:
The programmer/controller will not operate the unit's heaters (to change or maintain a temperature) unless there is a program
running or a single setpoint value is selected in the Local Mode.
Figure
.
Common Program Profile
Содержание Mini-Mite TF550 A
Страница 24: ...21 WIRING DIAGRAM 304242101...