Mini-Mite™ Tube Furnace
UP150 Controller Operation
9
7.3.2 Entering Program Parameters
The first display is the Start Set Point parameter, shown as
"SSP" in the upper display. The value assigned to SSP is
usually the current room temperature, 25°C.
On the next page is an illustration of the program profile and
a table of the parameters entered.
Basic Ramp and Dwell Parameters:
1.
Use the arrow buttons to select "25" in the lower display,
then press and release the 'SET/ENT' button twice to
enter this new value and to advance to the "StC" display. If
the value for "SSP" is correct and does not need to be
changed, press and release the 'SET/ENT' button once to
advance to the "StC" display.
2.
Next is the Start Code parameter, shown as "StC" in the
upper display. The value assigned to StC is usually."0".
This will instruct the program to follow the Start Set Pomt.
Press the SET/ENT button to advance to the next display.
3.
The next parameter, "SPI", is the first setpoint value that is
desired in the chamber and is normally a ramp segment.
Select this target temperature setpoint value with the arrow
buttons then press and release the 'SET/ENT' button twice
to enter this value and to advance to the "tMI" display. If
the value for "SPl" is correct and will not be changed, press
and release the 'SET/ENT' button once to advance to the
"tMI" display.
4.
The next parameter, "tMl", represents the first time period
for the unit to reach the target temperature setpoint
selected in "SPI". This selection can be a value ranging
from 0.00 to 99.59, which represents hours and minutes.
Select this time value with the arrow buttons and enter it by
pressing and releasing the 'SET/ENT' button twice.
5.
Press and release the 'SET/ENT' button to advance to the
next display of "SP2", this is normally the dwell segment.
Select the same target setpoint temperature value as "SPI"
with the arrow buttons. Press and release the 'SET/ENT'
button twice to enter this value and to advance to the next
display.
6.
The next parameter, "tM2", represents the second time
period used to maintain or dwell at the target setpoint
selected in "SP2". This selection can be a value ranging
from 0.00 to 99.59, which represents hours and minutes.
Select this time value with the arrow buttons and enter it by
pressing and releasing the 'SET/ENT' button twice.
7.
Next, "SP3" is the third setpoint value desired in the
chamber. Select this target temperature setpoint with the
arrow buttons and press and release the 'SET/ENT' button
twice to enter this value and to advance to the "tM3"
display. If this value is correct and not changed, press and
release the 'SET/ENT' button once to advance to the "tM3"
display.
8.
"tM3" represents the third time period for the unit to reach
the target setpoint selected in "SP3". This selection can be
a value ranging from 0.00 to 99.59, which represents hours
and minutes. Select this value with the arrow buttons and
enter it by pressing and releasing the ‘SET/ENT’ button.
9.
The next parameter, "SP4" is normally the dwell segment.
Select the same target temperature as "SP3" with the arrow
buttons, then press and release the 'SET/ENT' button twice
to enter this new value and to advance to the next display.
Additional Program Parameters"
10.
The next parameter displayed is "tM4". Select a lower
display value of "oFF" with the ‘arrow’ buttons, then press
and release the 'SET/ENT' button twice to enter this value
change and advance to the next display. ·
11.
The next display shows "EVl" in the upper display. The
lower value should always be "0" (zero). Press and release
the 'SET/ENT' button once to go to the next display.
12.
"ALl" should always have a lower value of "9". Press and
release the 'SET/ENT' button once to advance to the next.
13.
The next parameter, "AI", is used to select the high
temperature alarm trip setpoint. Use the ‘arrow’ buttons to
select a value 10°C (or 20°F) HIGHER than the highest
target setpoint to be used. Select the High Temperature
Alarm value with the arrow buttons then press and release
the ‘SET/ENT’ button twice to enter this new value and to
advance to the "HYl" display. If the value for "AI" is correct
and not changed, press and release the 'SET/ENT' button
once to advance to the "HYl" display.
14.
"HYl" is used to select the amount of temperature change
below the high temperature alarm setpoint where the alarm
relay will reset. This value is usually "I". Select "I" with the
arrow buttons and press the 'SET/ENT' button six times to
enter the correct value and advance to the "JC" display. Or
if the value is correct, press the 'SET/ENT' button five times
to advance to the "JC" display.
15.
For the parameter displayed as "JC", select "I" with the
arrow buttons, then press and release the 'SET/ENT' button
twice to display "WTZ". Selecting the value of"I" will cause
the program to hold the setpoint at this last segment. A
value of "0" would cause the program to reset and stop
running the program and stop the power to the heaters. A
value of "2" will cause the program to repeat 'continuously'.
16.
When the display shows "WTZ", select a lower display
value of"oFF" with the ‘arrow’ buttons. Press and HOLD the
‘SET/ENT’ button for 3 seconds to return to the Reset or
standby display.
This concludes the steps required to enter a typical ramp-and
dwell program. On the next page is an illustration of the
program profile and a table of the parameters entered.
Содержание Mini-Mite TF550 A
Страница 24: ...21 WIRING DIAGRAM 304242101...