background image

, a product of Lilliput

VDS6000 Series PC Oscilloscopes

(VDS6102/A, VDS6152/A)

User Manual

www.owon.com.cn

Содержание OWON VDS6000 Series

Страница 1: ...a product of Lilliput VDS6000 Series PC Oscilloscopes VDS6102 A VDS6152 A User Manual www owon com cn...

Страница 2: ...in this user manual supersedes that in all previously published material Specifications change privileges reserved OWON is the registered trademark of Lilliput Contacting Lilliput Fujian Lilliput Opto...

Страница 3: ...vice This warranty shall not apply to any defect damage caused by improper operation or improper inadequate maintenance towards OWON product Lilliput shall not be obligated to under this warranty a re...

Страница 4: ...e vertical system from PC software iii how to set the horizontal system from PC software iv how to set the trigger system from PC software v how to set channel from PC software vi how to use automatic...

Страница 5: ...g value of all terminals been well followed so as to avoid any potential short circuit or electric shock iv NO direct body touch with any naked conductor of device when working the device Naked conduc...

Страница 6: ...ted facility damage Terms on the device It covers Danger It means the operation may result in the immediate body injury Warning It means the operation may result in potential body injury Caution It in...

Страница 7: ...When working the device to avoid short circuit the ground of two probes are NOT allowed to connect to 2 different non isolated DC level The illustration of the device s built in ground wire connection...

Страница 8: ...II environment DO NOT get the 40 V input voltage from earth surface through any non isolated input DO NOT get the 40 V input voltage of dropout voltage through any non isolated input v DO NOT introdu...

Страница 9: ...fective Hard Disk Space 1GB Recommended System Requirement CPU Pentium Dual Core 2 4 GHz Internal Memory 2GB Effective Hard Disk Space 2GB Other Requirement Operating System Windows 10 or Windows 8 7...

Страница 10: ...when the device powered by PC through USB connection cable via USB communication interface of PC with PC powered by AC power source without adapter connection the input current should reach 1 5A or ab...

Страница 11: ...ware normally and smoothly NI VISA driver is a must Note On condition that NI VISA driver or similar VISA driver already been well installed onto the target PC step ii could be skipped i To download t...

Страница 12: ...via Next to enter into Select the installation directories window after locating installation path click Next...

Страница 13: ...then Next following window comes untick the box then Next...

Страница 14: ...Agreement s to continue Next Another Next based upon the step done last Next eventually press Restart until PC restarts the full NI VISA driver will be installed successfully iii to run PC Software Vi...

Страница 15: ...atus indicator lights red for seconds Via USB connection cable type C through matching communication interface to connect the device with PC When the status indicator lights green PC software detects...

Страница 16: ...p please refer to xii how to use main action button 8 Single Trigger please refer to xii how to use main action button 9 Back to Home of Function Menu 10 Hide the current menu 11 Voltage Range measure...

Страница 17: ...how to use automatic measurement 20 Cursor Measurement extension window please refer to ix how to use cursor measurement 21 Yellow Pointer to show the grounding base point zero point of Channel 2 prov...

Страница 18: ...error occurred ReSyncing re synchronize with the device AutoSet in process of auto setting Keyboard Shortcuts F5 Run Stop Ctrl Enter Auto Set Q 1 division less from Channel 1 s voltage division A 1 d...

Страница 19: ...setting from function menu will keep valid until new setting change introduces Caution The default probe compensation setting into PC software reads x10 before working the probe with the device makin...

Страница 20: ...arther slider from the center of control bar the faster the vertical position changes Another solution is to drag the zero point pointer upwards or downwards as item 22 and 21 goes under VII Operation...

Страница 21: ...izontal position of the signal as item 3 goes under VII Operation Interface of PC Software Keyboard Shortcuts 1 division less from time base 1 division more from time base iv how to set the trigger sy...

Страница 22: ...rigger option here indicate single trigger it is the one to use one trigger signal to capture data from two channels simultaneously the operation steps i choose CH1 or CH2 from Source ii next Mode 4 o...

Страница 23: ...ptions Auto in this trigger mode the device could capture the signal without set trigger condition Normal in this trigger mode only capture the signal with set trigger condition met Single in this tri...

Страница 24: ...ides the trigger time and the unusual pulse could be captured through setting special pulse trigger condition ii pick up suitable pulse Condition iii set pulse time iv set Trigger level v set Trig Mod...

Страница 25: ...oise of displayed waveform Fullband restore the channel bandwidth to full bandwidth to set channel coupling DC when this option chosen both direct and alternating components of the input signal could...

Страница 26: ...p Vmax Vmin Vtop Vbase Vamp Vavg Vrms Overshoot Preshoot Frequency abbreviated as Freq Period Rise Time Fall Time Delay A B Delay A B Width Width Duty Duty The Measure option covers several useful fun...

Страница 27: ...assist better understanding different measurement options Definitions about these options Vpp voltage between upper peak and lower peak from measured signal Vmax voltage between upper peak and ground...

Страница 28: ...tter understanding different measurement options Definitions about these options Rise Time the time span for signal amplitude rise from 10 to 90 judged by the rising edge of its first pulse Fall Time...

Страница 29: ...ignal noise Average this sampling mode is to reduce random irrelevant signal noise with set range from 1 to 128 1 the set no 128 Following illustration is signal output under normal Sampling mode no i...

Страница 30: ...er Average mode with average number set in 16 from here the interrupting signal noise been reduced to certain degree vertical resolution option this option is available only in VDS6102A VDS6152A Via P...

Страница 31: ...lable from the device with 4 basic waves output sine square ramp and pulse through MULTI communication interface From Function menu via enter into Function Generator option i choose preferred wave typ...

Страница 32: ...s indicating Cursor 1 and Cursor 2 When moving mouse pointer through Cursor 1 or Cursor 2 it shapes drag leftwards or rightwards to adjust the measurement range between Cursor 1 and Cursor 2 The curso...

Страница 33: ...and Cursor 2 The cursor measurement extension window located at the bottom left of main display area tells the current position of Cursor 1 Cursor 2 the absolute value of the voltage amplitude differ...

Страница 34: ...econd channel in Y axis Note When the device works in XY Mode the record length been restricted at 1K points The record length setting changes to 1K points automatically Persistence When Persistence c...

Страница 35: ...her Black or Blue After different color chosen from the current one Restart Effect option comes press Restart after the option the PC software will restart in newly set operation interface skin color...

Страница 36: ...tion Before doing this operation remove off all device connected items including but not limited to working probes leads clamps then press Self Cal until the full process ends Default Mouse click Defa...

Страница 37: ...usted to the proper division Horizontal Level middle Horizontal Sale been adjusted to the proper division Sampling Run Stop Status Run Acquisition Mode current Display Display Format YT Run Stop To st...

Страница 38: ...How to Communicate Device with PC press Network button under Utility option from function menu b mouse click LAN Setup to to enter into MachineNetSetting IP address total in 4 sections to set the fir...

Страница 39: ...through AC adapter or through USB connection cable the device been powered with indicator lights red for seconds b LAN connection To communicate device with target PC through LAN connection via respe...

Страница 40: ...via Devices and Interfaces then Network Devices enter into Add Network Device e press to dot tick Auto detect of LAN Instrument then Next f to following window...

Страница 41: ...ected on local subnet to choose TCPIP0 the previous IP connection just added inst0 INSTR then mouse click Finish g back to PC software to press TCPIP0 the previous IP connection just added inst0 INSTR...

Страница 42: ...unication interface after the communication successful please refer to iv to communicate device with PC under V How to Communicate Device with PC press Network button under Utility option from functio...

Страница 43: ...through AC adapter or through USB connection cable the device been powered with indicator lights red for seconds b LAN connection To communicate device with target PC through LAN connection via respe...

Страница 44: ...via Devices and Interfaces then Network Devices enter into Add Network Device e press to dot tick Auto detect of LAN Instrument then Next f to following window...

Страница 45: ...ected on local subnet to choose TCPIP0 the previous IP connection just added inst0 INSTR then mouse click Finish g back to PC software to press TCPIP0 the previous IP connection just added inst0 INSTR...

Страница 46: ...mouse click WiFi Setup to WiFiSetting choose AP as Connect Type to set SSID and Password IP here fill 192 168 100 1 Port here use 8866 could be any value ranging from 0 till 65536 c press OK to confir...

Страница 47: ...d following page comes via Devices and Interfaces then Network Devices enter into Add Network Device e press to dot tick Auto detect of LAN Instrument then Next...

Страница 48: ...ument s detected on local subnet to choose TCPIP0 the previous IP connection just added inst0 INSTR then mouse click Finish g back to PC software to press TCPIP0 the previous IP connection just added...

Страница 49: ...sful please refer to iv to communicate device with PC under V How to Communicate Device with PC press Network button under Utility option from function menu b in Network mouse click WiFi Setup to WiFi...

Страница 50: ...h WiFi hotspot set onto device as per correct SSID and Password Note Provided not possible to auto switch between WiFi and LAN connection onto target PC disable LAN Dot tick Obtain an IP address autom...

Страница 51: ...dow dot tick Select instrument s detected on local subnet to choose TCPIP0 the previous IP connection just added inst0 INSTR then mouse click Finish g back to PC software to press TCPIP0 the previous...

Страница 52: ......

Страница 53: ...t mode 20 MHz VDS6152A 8 bit mode 150 MHz 12 bit mode 100 MHz 14 bit mode 20 MHz Vertical Resolution A D VDS6102 VDS6152 8 bits VDS6102A VDS6152A 8 bits 12 bits 14 bits Channel Q nty 2 Acquisition Mod...

Страница 54: ...C 100MHz Single 1 interval time 25ppm x reading 0 6ns Average 16 1 interval time 25ppm x reading 0 4ns Vertical System Sensitivity 2 mV div 5 V div Displacement 2 V 2 mV div 50 mV div 20 V 100 mV div...

Страница 55: ...om the screen center Trigger Level Accuracy typical Internal 0 3 division Trigger Displacement on the basis of record length and time base Trigger Hold off Range 100ns 10s Edge Trigger Slope rising fa...

Страница 56: ...tion 10 mVpp General Part Communication Interface USB Device USB Host hi speed USB 2 0 LAN 10 100Mbits Programming Language SCPI Compatibility USBTMC LXI SOCKET Power Power Source 5V 15V DC 1 2A Power...

Страница 57: ...ies 2 x passive probe 1 x USB connection cable type C 1 x hard copy quick guide 1 x AC DC adapter 1 x BNC Q9 cable 1 x PC software user manual CD Optional Accessories 1 x Wi Fi module 1 x soft case No...

Страница 58: ...lity connected making sure probe not working with device outer facility ii Use non electrostatic soft cloth to remove the surface dust iii For further surface cleaning a wet but no water dripping soft...

Отзывы: