2.4 Trouble Shooting Viking™ XL mobile lift, 2040043
The charger doesn’t work.
The lift is stuck in the high position.
If you hear unusual sound from the lift.
1. Check that the charger cables are connected correctly.
2. Make sure that the battery is properly seated in the control box.
3. Try an alternative mains outlet.
4. If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop button has not been
activated (shall not be pressed in).
2. Make sure that the battery is properly seated in the control box.
3. Check the battery capacity.
Check if the Li-ION battery has been set in to sleep mode,
see chapter "Operation".
4. Check that the handcontrol cable is connected correctly.
5. Electrical emergency lowering, use the operation panel to
lower the patient onto a firm surface, see chapter; Operation.
6. Use the mechanical emergency lowering device to lower the
patient onto a firm surface, see chapter; Operation.
7. If the problem persists, please contact Hill-Rom.
Contact Hill-Rom.
The lift do not work up/down
with Hand Control.
The base-width adjustment
doesn’t work (in/out) with the
Hand Control.
1.
Make sure that the emergency stop button has not been
activated (shall not be pressed in).
2. Check the battery capacity.
Check if the Li-ION battery has been set in to sleep mode,
see chapter "Operation".
3. Make sure that the battery is properly seated in the control box.
4. Check that the charger cable is not connected to an electric outlet.
5. Check that the handcontrol cable is correctly connected to
the control box.
6. Check that the lift arm actuator cable is correctly connected to
the control box.
7. Check that the base width actuator cable is correctly connected to
the control box.
8. If the problem persists, please contact Hill-Rom.