![Liko Uno 102 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/liko/uno-102/uno-102_instructions-for-use-manual_1916145009.webp)
9
7EN150104 Rev. 8
• 2020
www.hillrom.com
Guidance and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
”Essential performance according to the manufacturer: The product shall not move unintentionally while being submitted
to disturbances.”
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST Level
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
�)
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c)
+/- 5 kHz deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE Band 13, 17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800 - 960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700 -
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band 1, 3, 4, 25
UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400 -
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5100 -
5800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
NOTE If cecessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or
ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulatted using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.