Liko Uno 102 Скачать руководство пользователя страница 5

5

7EN150104 Rev. 8  

•  2020

www.hillrom.com  

Definitions

1.  Lift arm

2.  Flexlink

3.  Latches

4.  Sling bar

5.  Outer tube

6.  Motor for lift arm

7

.  Product label

8.  Front wheels

 

9

.  Base

10

.  Rear wheels with brakes

1 1.  Motor for base-width adjustment

12

.  Locking handle 

13

.  Cable for hand control

1

4.  Control box

1

5.  Battery

16

.  Hand control

17

.  Lift mast

1

8.   Holder for Quick Reference Guide and colour 

codes for sling sizes

19

.  Emergency lowering (mechanical)

20.  Emergency stop

21.  Emergency lowering, electrical

22.  Charge indicator (Charge = charges)

23.  Charge indicator (On = charger connected)

Technical Data

The device is intended for use indoors.
Type B, in accordance with the electrical shock protection 

class.
Class II equipment.

Forw

ard dir

ection

Maximum load:

175 kg (385 lbs)

  

Material:

Powder-painted steel

Weight:

Total:

 

42,7  kg (94.1 lbs)

Heaviest removable part: 

22,3 kg (49.2 lbs)

Wheels:

Front: 75 mm (2.9 in) twin wheel

Rear: 75 mm (2.9 in) individual wheels fitted 

with brakes

Turning 

Diameter:

1380 mm (54 in)

Emergency 

lowering device:

 

Mechanical and electrical

Lift Interval:

1270

 mm (50 in)

Lifting speed:

30 mm/sec (1.18 in/sec) without load

Sound level:

39 

dB(A)

 

Protection class:

IP X4

Operating Forces 

of Controls:

Buttons on hand control: 

2.4 N

Electrical  

data: 

 

24 V

Intermittent 

power:

Int. Op 10/90, active operation max 2 min. 

Only 10% of a given length of time may be 

active, but no more than 2 min.

Batteries:

2 x 12 V 2.9 Ah valve-regulated lead-acid 

gel-type batteries. New batteries are provided by 

the supplier.

Battery charger:

 Built-in, 100-240 V,AC, 50-60 Hz, 

 

max 400 mA.

Lift motor:

24 V 6 A, permanent magnetic motor with 

mechanical safety mechanism, safety nut 

and outer tube.

Motor for 

base-width 

adjustment:

Surrounding 

functional 

environment:

 

 

24 V 3.5 A, permanent magnetic motor

Temperature: +10°C to +50°C, (50° F to 122° F)

 

Humidity: 20% to 90% at 30°C non-condensing, 

Atmospheric pressure: 700-1060hPa.

21
22

23

14

20

9  

3

5

8

1

17

7

19  

2

10

13

14

16

15

18 

6  

12

11

  

Содержание Uno 102

Страница 1: ...utmost importance for function and safety when using the lift In this document the person being lifted is referred to as the patient and the person helping them is referred to as the caregiver Product...

Страница 2: ...Instructions 4 Definitions 5 Technical Data 5 Dimensions 6 EMC Table 7 Assembly 10 Operation 12 Charging the Batteries 14 Maximum Load 15 Recommended Lifting Accessories 15 Simple Troubleshooting 17...

Страница 3: ...line underneath the symbol indicate this product have been placed on the market after 2005 UL Recognized Component Mark for Canada and the United States EFUP Environmental Friendly Usage Period years...

Страница 4: ...g accessory hangs vertically and can move freely the lifting accessory is selected appropriately in terms of type size material and design with regard to the patient s needs the lifting accessory is c...

Страница 5: ...ont 75 mm 2 9 in twin wheel Rear 75 mm 2 9 in individual wheels fitted with brakes Turning Diameter 1380 mm 54 in Emergency lowering device Mechanical and electrical Lift Interval 1270 mm 50 in Liftin...

Страница 6: ...urement Table Measurements in mm Model A Max A Min B B1 B2 B3 C D D1 D2 E F F1 L Max L Min M N O max min max min Uno 102 79 3 77 4 75 8 56 5 54 5 53 0 49 4 36 2 28 7 23 2 42 9 27 2 38 6 22 8 37 4 1 8...

Страница 7: ...test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 2 kV 4 kV 8 kV 15 kV air 8 kV contact 2 kV 4 kV 8 kV 15 kV air Fl...

Страница 8: ...is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each freque...

Страница 9: ...745 780 810 800 960 GSM 800 900 TETRA 800 IDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulatio...

Страница 10: ...sired position with the locking handles provided After securing the mast use Pull out rotate to set the locking handles in a downwards position see illustration 2 The lifting height can be set to thre...

Страница 11: ...properly 7 Hang the hand control on the handle 8 Reset the emergency stop by turning the button clockwise ON CHARGE BATTERY 9 A Connect the extension cable for the charging cable to the control box B...

Страница 12: ...yckt cirkel 46 mm F rgkoder Cirkel R d PMS 185 C Ilustration 100 Svart Locking the Wheels The rear wheels can be locked to prevent rotating and turning The locking unlocking of the wheels is done with...

Страница 13: ...he floor Lift correctly Before each lift make sure that the Sling loops at opposite sides of the Sling are at the same height all the Sling loops are fastened securely in to the Slingbar hooks the Sli...

Страница 14: ...lights up green when the charger is connected to mains 2 CHARGE lights up yellow constantly during charging and will turn off when charging is completed Charging with the control box internal charger...

Страница 15: ...xample a Uno 102 mobile lift which is approved for 175 kg 385 lbs can be equipped with a lifting accessory which is approved for 300 kg 660 lbs In this case the maximum load of 175 kg 385 lbs applies...

Страница 16: ...l charger and battery Prod No 2107103 Quick Release Hook Liko Quick Release Hooks is a system for quick change of lifting accessories on Liko s mobile and stationary lifts Uno mobile lift must be equi...

Страница 17: ...Check the battery capacity 4 Check that the handcontrol cable is connected correctly 5 Electrical emergency lowering use the operation panel to lower the patient onto a firm surface see chapter Operat...

Страница 18: ...gs To help prevent injury and or equipment damage obey these warnings Warning The potential for electrical shock exists with electrical equipment Failure to follow facility protocol could cause death...

Страница 19: ...phor and zinc oxide creams can cause permanent stains Remove temporary stains by wiping vigorously with a lightly dampened wiping cloth Cleaning and Disinfection Overview Cleaning and disinfection are...

Страница 20: ...TE It is important to remove all visible soil from all areas before continuing to remove non visible soil With a new wiping cloth soaked in an approved cleaner disinfectant use firm pressure to wipe a...

Страница 21: ...Golvo and ceiling lifts Alcohol Isopropyl alcohol 70 5 0 7 0 Viraguard Veridien Hand controls for all lifts Quaternary ammonium n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl ethyl...

Страница 22: ...product has an expected service life of 10 years when correctly handled serviced and inspected in accordance with Liko s instructions Transport and Storage During transport or if the lift is not goin...

Отзывы: