Service Manual Golvo™ 8000/8008
Mobile Lift
• 3EN400404 Rev. 6 19
www.liko.com
l
www.hill-rom.com
6.1 Service, Control Unit CBL - Parameter Code Key
Control Unit CBL
Parameter:
Information:
TAAxT
[Ah]
Totally Accumulated Amphours (current x time)
TAAxTR
[Ah]
Totally Accumulated Amphours, resetable (current x time)
Can be reset by the manufacturer.
PAAxT
[Ah]
Periodically Accumulated Amphours (current x time)
Calculated continuously from the moment the lift is used until the first
service.
PAKT
[Months]
Periodically Accumulated Calendar time.
Calculated between services.
TADT
[Hours]
Totally accumulated drive time.
AAS
[Starts]
Accumulated starts.
The number of lifting motions counted from the moment the lift was
used the first time.
ASERV
[Services]
Accumulated services
Number of performed service events made using LSS or LDS.
IMED
[A]
Average power consumption during TADT.
Overall accumulated operating time.
ALA
[Charges]
Accumulatet number of started chargings.
AALA
[Charges]
Accumulated number of completed chargings
AOLA
[Charges]
Accumulated numbers of not completed charging.
Counts each disrupted charging event.
AABT
[Removals]
Accumulated number of battery replacements.
Counts each battery replacement where a used battery is replaced.
AEUP
[Endpos. up]
Accumulated number of end positions during lifting motion up.
AFSERV
[Services]
Accumulated number of reset Service
Number of “Clear Service” performed manually.
AEDWN
[Endpos. down]
Accumulated number of end positions during lifting motion down.
SFSTAT
N/A (This funktion is not in use)
IDNR
Identity number of the control box.
AOLUP
[Overloads]
Accumulated number of stops due to overload.
PADT
[Hours]
Periodically accumulated operating time.
IMED 20
[A]
Average current consumption. Recounted for each lifting motion (up)
which lasts more than 20 seconds.
ELT
[Ah]
TAAxTR limit (constant) for ELT indication.
ELTCNT
[Commands]
Number of ELT resets (only performed at replacement of actuator).
CLIM_1
[A]
Current limit for lifting motion (up).
CLCT
[ms]
Maximum allowed time to override CLIM_1 before cut-off
CLIM_2
[A]
Current limit for widening motion (base-width adjustment).
SINTT
[Months]
Indicates service intervals in calendar months.
SINTAS
[Ah]
AxT limit for service event. (current x time)
CYCLET
[Hours]
Cycle time is a constant for a medium driving cycle.
PAOLUP
[Overloads]
Periodic accumulated number of stops due to overload.
CBL PARAMETER Code Key
Liko Diagnostic System™
(operating data) is stored in Control Unit CBL
•
The information listed above is stored in the Control Unit CBL.
•
Replacing the Control Unit CBL without using the LSS may cause some data loss as this data is recordable by LSS and a
computer. When data is stored in Liko™ database it can be retrieved if needed.
•
Resetting the Service Interval using the LSS, LDS or Manually, will reset the Accumulated Amp-hours (PAAxT),
Accumulated Time (PAKT), Accumulated Operating Time (PADT) and Accumulated number of stops due to overload
(PAOLUP).