Service Manual Golvo™ 8000/8008
Mobile Lift
• 3EN400404 Rev. 6 9
www.liko.com
l
www.hill-rom.com
15
w w w . l i k o . com
Golvo • 7EN140107-01
The lift cannot be operated
with the hand control.
1. Check that the emergency stop button has not been pressed.
2. Check the battery capacity.
3. Check that the charger cable is not connected to
an electric outlet.
4. Check that the hand control cable is correctly connected.
5. If the lift works via the operation panel, change the
Hand Control.
6. If the problem persists, please contact Liko/Hill-Rom.
The lift does not work up/down
with the operation panel.
The base-width adjustment doesn’t
work (in/out) with the operation
panel.
1. Check that the emergency stop button has not been
pressed.
2. Check that the cables to the control box are connected
correctly.
3. Check that the charger cable is not connected
to an electric outlet.
4. Check the battery capacity.
5. If the problem persists, please contact Liko/Hill-Rom.
The charger doesn’t work.
1. Check that the emergency stop button has not been
pressed.
2. Check that the charger cables are connected correctly.
3. Make sure that the battery is properly attached.
4. If the problem persists, please contact Liko/Hill-Rom.
The lift is stuck in the high
position.
1. Check that the emergency stop button has not been
pressed.
2. Check the battery capacity.
3. Check that the hand control cable is connected correctly.
4. Electrical emergency lowering, use the operation panel
to lower the patient onto a firm surface.
5. Use the mechanical emergency lowering device to lower
the patient onto a firm surface.
6. If the problem persists, please contact Liko/Hill-Rom.
The lift does not reach maximum
lifting height.
1. Check that the lifting interval level is correctly set.
2.
If the problem persists, please contact Liko/Hill-Rom.
If any noises are heard.
Contact Liko/Hill-Rom.
Simple Troubleshooting
2.2 Troubleshooting, Golvo™ 8000/8008 mobile lift, with fixed hand control