10
www.hillrom.com
7EN120114 Rev. 10 • 2020
Electrical Emergency lowering
In the event that the hand control or electronics
malfunction, the lift can be lowered by pulling on
the emergency lowering device.
In the event of emergency lowering, always ensure the
patient is lowered into a bed, wheelchair or other suitable
place.
Emergency stop
To activate the emergency stop
: pull the red emergency
stop string.
To reset the emergency stop
: turn the red emergency stop
button in the direction shown by the arrows.
The red button on the
lift unit
end cover is intended
for use if an emergency situation occurs. When the button
is pressed, contact between the motor and power source is
broken, which stops the lifting motion.
Installation of Latches
After installation, check that the
latch locks and runs freely in the
sling bar hook.
Lift correctly!
Before each lift, make sure that:
– the Sling loops at opposite sides of the Sling are at the same height
– all the Sling loops are fastened securely in to the Slingbar hooks
– the Slingbar is level during the lift, see Figure 1.
If Slingbar is not level (see Figure 2) or if the sling loops are incorrectly at-
tached to the slingbar (see Figure 3) lower the user to a firm surface and ad-
just according to the Instructions for use of Sling in use.
An improper lift can be uncomfortable for the user and cause damage to
the lift equipment! (see Figure 2 and figure 3).
Figure 1.
Figure 2.
Figure 3.