Programming
248083201-1 Rev 6
4-3
DIP Switch Settings and Meanings
1000:1000 ==> 6 seconds, 1000 counts
1000:0100 ==> 6 seconds, 10,000 counts
1000:0010 ==> 6 seconds, 100,000 counts
0100:1000 ==> 60 seconds, 1000 counts
0100:0100 ==> 60 seconds, 10,000 counts
0100:0010 ==> 60 seconds, 100,000 counts
0010:1000 ==> 600 seconds, 1000 counts
0010:0100 ==> 600 seconds, 10,000 counts
0010:0010 ==> 600 seconds, 100,000 counts
If the user chooses DIP switch combinations other than the ones
specified above, the unit will default to a sample time of 300 seconds,
Range1 and Range2 to 100,000 counts.
Procedure to Set DIP Switches
1. Remove power from the instrument.
2. Set the DIP Switches to the desired Time and Range, using the
information above.
3. Apply power to the instrument.
Example Startup Echo, Reading DIP Switches:
Lighthouse Remote Instrument (4-20mA):
- reading dip switch:
Sample Time = 60 secs
Range1= 10000.
Range2= 1000.
Alarm Threshold = 750 counts
Suppress Alarms = 2
Service High = 3500 mV
Service Low = 125 mV
Содержание REMOTE 2 Series
Страница 1: ......
Страница 2: ...00...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual t iv 248083201 1 Rev 6...
Страница 12: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual ii 248083201 1 Rev 6...
Страница 18: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual 2 4 248083201 1 Rev 6...
Страница 46: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual 5 4 248083201 1 Rev 6...
Страница 50: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual A 2 248083201 1 Rev 6...
Страница 53: ...Index 248083201 1 Rev 6 I 3 Weight 2 3 Z Zero Count Check Troubleshooting 6 2 Zero Count Level 2 3...
Страница 54: ...Lighthouse REMOTE 2012 3012 5012 3102 5102 Operating Manual I 4 248083201 1 Rev 6...
Страница 55: ...00...
Страница 56: ......