Lighthouse REMOTE 1100/1104 Operating Manual
5-2
248083363-1 Rev 4
3.
Configure the unit via the user interface to sample for 30 minutes.
4.
Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles
that might be inside it.
5.
Configure the unit via the user interface to sample for 5 minutes
and set a 10 second hold.
6.
Set Cycles to 10 so the instrument will take 10 five-minute
samples.
7.
If an average of more than one count per five minute sample is
reported, reset the instrument to sample for 30 minutes again to
purge it, then repeat the Purge Test again.
8.
After the instrument has met the requirement of the Purge test,
remove the filter, cap the inlet for moving and return the
instrument to its normal location and operating status.
9.
If the instrument still fails the Purge Test, contact Lighthouse
Tech Support for assistance.
Содержание REMOTE 1100
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Страница 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Страница 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Страница 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Страница 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Страница 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Страница 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Страница 69: ...00...
Страница 70: ......