Lighthouse REMOTE 1100/1104 Operating Manual
4-6
248083363-1 Rev 4
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 4-5 Instrument Parameters
Register
Data Type
Description
40026
unsigned integer
Location number; Specifies location of
Particle Counter.
40029
unsigned integer
Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 359,999, which equals 99h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
359,999, which equals 99h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Содержание REMOTE 1100
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Страница 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Страница 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Страница 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Страница 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Страница 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Страница 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Страница 69: ...00...
Страница 70: ......