18
How to do a control solution test:
Step 1: Press On/Off Button. Insert
Test Strip.
The code number that
appears on the meter
display should match
the code number on
the test strip vial.
When
insert code10
(Example)
s t r i p
a p p e a r s , slide a
ONE TOUCH
™
Test Strip into the
test strip holder, notched end first,
with the test spot
facing up. Make sure
you push the test strip
in all the way until it
stops.
w a i t
appears
for a moment, then
apply sample
appears for five minutes or until you
apply a drop of control solution to
the test spot on the test strip.
M
M
C
M
M
C
(Example)