MESSAGE
PROBLEM
WHAT TO DO
clean
test
area
1. There is dirt, blood, or
lint on the test area.
2. Your hand or an object
covered the test area
while the meter was
turned on.
3. The meter was used in
very bright light.
The test strip was
inserted before the
word
insert
appeared
on the display.
1. Clean the test area and
test strip holder according
to instructions.
2. Repeat the test. Keep the
meter’s test area clear.
3. Move the meter away
from the light source.
Repeat the test with a
new test strip.
Repeat the test with a new
test strip. Wait for the word
insert
before inserting a
test strip.
remove
strip
52