EVOS
®
FL and EVOS
®
FL Color User Guide
37
Changing LED Light Cubes
Install Light Cube
Optional LED light cubes are available; contact Technical Support (page 54) for
details. Each LED light cube is coded so the EVOS
®
FL Imaging System will
automatically recognize it in any position.
WARNING! UV LIGHT HAZARD!
This instrument uses a Class 3B
ultraviolet LED for the DAPI channel. Before changing LED light cubes,
power off the instrument.
1.
Turn the power switch to the “
” (OFF) position before removing or
changing light cubes.
Figure 33
Power switch
2.
Move the stage back to allow access to the light cube access cover, centered
under the back of the stage. (Figure 34)
Figure 34
Light Cube Access Cover (under stage)
3.
Loosen the thumbscrew to remove the light cube access cover and remove the
light cube tool under it. (Figure 35)
Figure 35
Light Cube Tool (under access cover)