where
V
c
and
V
co
are the raw detector signals for sample and reference,
Z
c
is the
CO
2
zero parameter, and
X
wc
is a cross sensitivity parameter for the effect of water
vapor on CO
2
. Its value is reported on the calibration sheet as
XS=
.
The empirical pressure correction function
g
c
() depends on CO
2
absorptance and
pressure:
When
P
=
P
o
,
g
c
() = 1.
When
P
<
P
o
A-11
where
a
= 1.10158,
b
= -6.1217E-3,
c
= -0.266278,
d
= 3.69895, and
z
is the asymp-
totic value of absorptance, obtained from the calibration coefficients (equation
A-15).
A-12
When
P
>
P
o
A-13
where
X
,
A
, and
B
are computed as in equation A-11. CO
2
concentration
C
(µmol
mol
-1
) is computed from
Appendix A. Equation Summary
A-3
Appendix A. Equation Summary