Section 1. General description
Installing device drivers
If you are using Windows 7 and 8, you'll need to install device drivers. After con-
necting the LI-830 or LI-850 to your computer with the USB cable, follow these
steps:
1
Download the drivers and unzip the file.
Drivers are available at
https://www.licor.com/documents/pamrygkdwsl8c61wj6y815gnf4tukh3t
2
Press the Windows key (
), type
Device Manager
, and press
Enter
to open your
device manager.
3
Under
Other Devices
, select the LI-830/850.
4
Right-click LI-830/850 and select
Update Driver Software
.
5
Browse and select the folder with the driver files.
6
Click
Next
to finish installing the drivers.
When they are installed, you should see
LI-COR 8x0
under
Ports (COM & LPT)
.
Storing the gas analyzer
When you are done using the gas analyzer, follow these practices to be sure that it
is ready to use the next time you need it. The instrument can be stored in non-con-
densing conditions. If the instrument is stored in high humidity or moderate
humidity for a long time while powered off, some optical components may tem-
porarily become saturated with humidity. If this occurs, the instument may need to
run for several days before readings return to normal. When storing the instru-
ment.
l
Install the covers over the air inlet and outlet. This will keep dust, bugs, and
other contaminants out of the optical cell.
l
Disconnect the power supply and power wires.
l
If possible, store the instrument in an air conditioned environment.
1-8
Using the LI-830 and LI-850 Gas Analyzers