6.
Locate the new wiring harness (p/n 9981-176). Route
the harness down through hole shown in Figure 5
(solenoids in photo removed for clarity); you can route
either end, as the connectors on both ends are the
same. Route the harness underneath the mounting
cage, and toward the access panel at the front of
instrument case.
NOTE
: Be careful not to use tools
that can damage the circuit board or other components
when routing the wire harness. Plug the connector
into the PUMP connector (Figure 5b). Replace the
access panel. Plug the other connector into the new
pump as shown in Figure 5c.
Note the orientation of
the red and black wires on the two connectors
.
7.
Install the new pump adapter plate (p/n 9881-204) at
the location of the pump mounting bracket removed in
Step 5. Use 2 of the 140-04315 cap screws and the
7/64” hex key to install the adapter plate, as shown in
Figure 6.
8.
Cut two 6” (15 cm) pieces of urethane tubing (p/n 222-
00303) from the length provided. Attach these two
pieces to the hose barbs shown in Figure 7 (you
removed two short lengths of tubing from these barbs
in Step 5).
Use 2 cap
screws to attach
the pump
adapter plate.
Plug the other
end of the wiring
harness onto the
connector on the
bottom of the
pump.
Attach two 6”
lengths of tubing
to these hose
barbs.
Route the wiring
harness down
through this hole
in the mounting
cage...
5a
...under the
mounting cage,
and onto the
circuit board
under the access
panel. Plug into
the Pump
connector.
b
6
7
Pump mounting
plate.
2
c