Troubleshooting
Soil Chamber Problems
Using the LI-6400
20-39
tor), put a pencil mark on the outside of the two halves so that it can be
reassembled correctly.
Soil Chamber Problems
Pump DoesnÕt Run
When configured for the 6400-09 Soil Flux Chamber, the pump turns on and
off automatically during the measurement cycles. If the sample cell CO
2
con-
centration is above the lower window value (target - delta), then pump should
come on when you start a measurement until
CO2S_µml
falls below that val-
ue. When
CO2S_µml
rises above the upper window value ( delta), the
pump should turn back on, unless the number of requested cycles has been
completed. If this is not happening, it could be for the following reasons:
1
OPEN 3.01 and no CO
2
mixer installed
There is a bug in OPEN 3.01 that affects units configured for soil flux mea-
surements that do not have a CO
2
mixer installed. Either upgrade to OPEN
3.2, or else do the following: 1) Press
K
in OPENÕs main screen, 2) At the ok
prompt, type
4000 nullonflow
and press
enter
, 3) Press
escape
to return to
the main screen.
2
Check the auxiliary operating parameters
There are 4 parameters (3 in OPEN 3.01) accessible in
f2
level 3
(
AUX OP Params
), that influence the pumpÕs behavior.
ÒExtra Draw Down
(ppm)Ó
can be used to intentionally overshoot the lower limit, providing extra
time for things to stabilize before the next measurement cycle.
ÒDead Time
(secs)Ó
prevents measurements from starting too soon after the pump turns
off.
ÒMin Measure Time (secs)Ó
is the minimum measurements time, and
prevents the pump from turning on again too soon.
ÒFlow during DrawDownÓ
(in version 3.2) lets you set the approximate flow rate when the pump is on.
75%
25%
Figure 20-11. Schematic representation of the LI-6400Õs flow divider,
that partitions flow to the sample and reference cells of the IRGA.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...