Sensor Calibration
11
Converting Calconstants to Multipliers
For LI-COR sensors that are not type 'SA' (i.e, type 'SB',
etc.), the calibration multiplier must be calculated from the
calibration constant given on the certificate of calibration.
EXAMPLE:
Calculate the multiplier for a LI-190SB
Quantum sensor with a calibration constant of 8.0
µ
Amps/1000
µ
mol s
-1
m
-2
.
Multiplier
=
−
( )
µ
(
)
µ
1 1000
8
s m
Amps
-1
-2
mol
= −
µ
µ
125 0
1
. mol s m
Amp
-1
-2
EXAMPLE:
Convert a calibration multiplier of 3.6
Klux/mAmp to footcandles (see Appendix B for conversion
factors).
Multiplier
(fc) =
3.6 Klux
Amp
lux
1 Klux
fc
1 lux
µ
×
×
1000
0 0929
.
=
µ
334 44
.
fc
Amp