ADVANCED MICRO CALIBRATION BATH TEMPERATURE
LCA 30/50
www.leyroinstruments.com
23
View/edit details:
-
Name: You can choose the necessary name.
-
Password: A password can be entered, so that it is only activated when that
password is entered.
-
Data logging: If activated, once the profile is started, registration is activated and if it
is deactivated, registration will not start when the profile is started and it will have to
be started manually if necessary
Export file:
In this option you can export the desired profile. To export it, an USB device must
be connected.
Delete profile:
This function serves, as its own name indicate, to eliminate the unnecessary
profile.
-
Create profile:
Use to create a profile
-
Name: You can choose the necessary name.
-
Password: A password can be entered, so that it is only activated when that
password is entered.
-
Data logging: If activated, once the profile is started, registration is activated and if it
is deactivated, registration will not start when the profile is started and it will have to
be started manually if necessary.
-
Go to profiles:
Allows you to view all created profiles.
-
Clear state profile:
This option allows you to delete the finished or stopped profile in the screen,
which m, in the screen shows, when a profile is initiated, the start profile, ,the remaining steps and the
estimated time and when the profile is finished or stopped manually, the name of the profile appears on
the screen. This function is use to erase that indication only when it is finished or stopped.
-
Cancel
Use to exit from the menu.
Observations:
The equipment has a manufacturer profile called calibration.
9. PT 100
Change from PT100 Int to PT100 Ext.
To change from internal to external probe it is necessary that the external probe is connected and press the
PT 100 Ext key, which will turn green, since by default the equipment works with the internal probe and with
the key in grey
Note:
If the equipment is started without the external probe connected, an error will appear, which will fire and
remain hidden, but the error will always appears, and it will be indicated by red flashes in the upper line. The
error will disappear when the external probe is connected and it is not necessary to turn off the equipment
To calibrate both, the internal and external probe, press the menu key, configuration, machine controls and
linearization.
Calibrate
To calibrate the internal probe, linearization 1 is chosen and linearization 2 is chosen for the external probe.
The function must be interpolated and the units must be in source.
To adjust both the internal and external probe by comparison with a standard, the same set point must be set
both in the equipment and in the linearization in the entry point 1. When the equipment has reached the
temperature, it is observed that it marks the probe to be calibrated and which marks the pattern and if the
probe to be calibrated is out of adjustment, the value that should mark at the exit point 1 will be introduced.
Содержание LCA 30
Страница 6: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 4 ...
Страница 30: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 28 ...
Страница 32: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 30 ...
Страница 34: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 32 ...
Страница 36: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 34 ...