![LEYBOLD THERMOVAC TTR 91 R Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/leybold/thermovac-ttr-91-r/thermovac-ttr-91-r_operating-manual_1890897006.webp)
6
300710168_002_C1 - 01/2018 - © Leybold
Technical data
Mechanical data
Dimensions
Refer to the next page
Mass
85 g
Internal volume of tube
5 cm
3
Enclosure rating
IP40
Performance, operating and storage conditions
Displayed pressure range
5×10
-4
to 1000 mbar
Accuracy
typically ± 15 % at < 100 mbar
Ambient temperature
Operation
5 to 60 °C
Storage
-30 to +70 °C
Bakeout
150 °C (with electronics housing removed)
Humidity
80 % RH up to 31 °C decreasing linearly to 50 % RH
at 40 °C and above
Maximum altitude
3000 m (indoor use only)
Maximum internal pressure
10 bar absolute (9 bar gauge)
Filament temperature
100 °C above ambient
Materials exposed to vacuum
Filament
Tungsten / Rhenium
Tube
Stainless Steel 316L & 304L
Filter
Stainless Steel
Other
Glass, Ni, NiFe, Stainless Steel
Electrical data
Supply voltage
nominal
15 to 30 V d.c.
minimum
13.5 V
maximum
32 V
Maximum power consumption
1 W
Max inrush current
150 mA
Electrical connector
FCC68 / RJ45 8-way
Pressure output
Sensor output
1.9 to 10.0 V
Sensor fail
0.5 V
Min load impedance
10 k
Ω
Max output current
1 mA
Set-point
Adjustment range
1.4 to 10.0 V
Hysteresis
500 mV
Max external load rating
30 V d.c., 100 mA
Gauge identification resistance
27 k
Ω
Содержание THERMOVAC TTR 91 R