Gas dependency
The TTR 91 R is calibrated for use in nitrogen, and will read correctly with dry air, oxygen and carbon monoxide. For
any other gas type a conversion is required in order to obtain the correct pressure reading. The graph below shows
the conversion for 6 common gases: nitrogen, argon, carbon dioxide, helium, krypton and neon.
Gas dependency of TTR 91 R
For pressures below 1 mbar a simple calibration factor can be used to correct for different gas types. Gas Calibration
Factors (GCFs) for common gases are shown below.
True pressure = GCF x indicated pressure
Gas calibration factors below 1 mbar
Gas
GCF
He
1.1
Ne
1.5
N
2
1.0
Ar
1.7
CO
2
1.0
Kr
2.6
He
Ne
N
2
CO
2
CO
2
Ar
Kr