Operation
Pressure measurement
When the TTR 91 R is connected to a power supply the status LED will turn amber for approximately 2 seconds.
The status LED will then turn green if the gauge is operating correctly or red if an error is detected. Refer to the fault
finding guide.
If the gauge is connected to a Leybold controller the display will indicate the measured pressure.
The TTR 91 R provides a voltage output which is a function of pressure. The output voltage scales with 1.296V dc /
decade.
Conversion formulae:
𝑃
𝑚𝑏𝑎𝑟
= 10
(𝑉−6.143)/1.286
𝑉 = 𝑙𝑜𝑔
10
(𝑃
𝑚𝑏𝑎𝑟
) × 1.286 + 6.143