300668736_002_C0 - 11/2017 - © Leybold
1
Replace the tip seals
Replace the tip seal as described below. Prepare the pump by running on
gas ballast control for one hour to remove any process gases.
1. Switch off the pump and isolate from the electrical supply.
2. Using a 4 mm Allen key, undo the two screws (Figure 1, item 1) which
secure the fan cowl and remove.
3. Disconnect the motor fan lead from the controller (figure 2, item 1).
Figure 2 - Motor fan lead connector
1
Figure 1 - Fan cowl removal
1