Maintenance
30
300692481_002_C0 – 08/2018 – © Leybold
5.4
Top up the end cover oil level
WARNING
Turn off the seal purge. Allow the pressure in the end covers to reach to
atmospheric pressure. To remove an oil filter plug, undo it slowly to allow the
pressure in the end cover to equalize to atmospheric pressure, then fully undo and
remove the oil filler plug. If the instruction are not followed, hot oil can get ejected
from the end cover and cause injury.
WARNING
Ensure that the correct grade of oil is used and that the oil levels in the RUVAC
WA 40000 pump are correct. If an incorrect oil is used or if an oil level is incorrect,
pump performance may be affected and the pump may be damaged.
1. If the pump is operating, shut the pump down (refer to
) and allow the pump to cool to
a safe temperature.
2. Carefully remove the oil filler plug (10) from the filler port on the top of the drive end cover (9):
refer to the Warning at the start of
3. Pour new oil of the correct type (see
) through the filler port and into the end cover
until the oil-level reaches the maximum level mark. If the oil level goes above the maximum level
mark, drain oil from the pump until the level is correct: refer to
4. Refit the oil filler plug (10) to the filler port on the top of the drive end cover (9).
5.5
Inspect the system installation
Use the following procedure to inspect the system connections:
•
Inspect all the system pipelines and connections and check that they are not damaged or
corroded and that they are sufficiently leak tight. Repair or replace any damaged or corroded
component and seal any leak found.
•
If drive/transmission system is installed, inspect the drive/transmission system and adjust,
repair or replace as necessary: refer to the manufacturer’s instructions supplied with the
drive/transmission system.
•
Inspect all of the cooling water pipelines and connections and check that they are not
damaged or corroded and that they do not leak. Repair or replace any damaged or corroded
component and seal any leak found.
•
Inspect all of the electrical cables; check that they are not damaged and have not
overheated. Replace any cables that are damaged or have overheated. Check that all the
electrical connections are secure; tighten any loose connections.
Содержание RUVAC WA 40000
Страница 2: ...This page has been intentionally left blank...
Страница 4: ...This page has been intentionally left blank...
Страница 8: ...This page has been intentionally left blank...
Страница 41: ...300692481_002_C0 08 2018 Leybold 41 Declaration of conformity...
Страница 42: ...42 300692481_002_C0 08 2018 Leybold Declaration of contamination...
Страница 43: ...300692481_002_C0 08 2018 Leybold 43...
Страница 44: ...44 300692481_002_C0 08 2018 Leybold...