8
Trouble Shooting
Symptom
Possible Cause/Remedy
No digital communication
-
Check electrical connections (3 wires from transmitter to
communication equipment)
-
Transmitter and communication equipment baud rate
not matched.
-
Incorrect or unstable pressure
value
-
Transducer not zeroed.
-
Contaminated sensor.
-
Corroded sensor.
-
Transmitter exposed or heating or cooling air stream.
Incorrect pressure value at low
pressure.
-
Transducer not zeroed.
-
Contaminated sensor.
-
Corroded sensor.
-
Transmitter exposed to heat or cooling air stream.
Incorrect pressure value at high
pressure.
-
Transducer not zeroed.
-
Contaminated sensor.
-
Corroded sensor.
-
Transmitter exposed to heat or cooling air stream.
Zero pushbutton does not work
-
Pressure higher than 20% of full-scale measurement
range.
-
Electronics fault.
-
Zero pushbutton is damaged (user induced).
No analog output
-
Power supply turned off.
-