62/102
17200234_002_C1/2016 - © Leybold
602494EA.DOCX
The Edit symbol is displayed in the «Nom», «Min» or
«Max» line. Press
to adjust a value.
to save the new setting and return to the
«Pressure» display.
to return to the «Pressure» display
without saving the new setting.
Selecting the setpoint source for the Pressure mode
The setpoint (nominal value) source setting made for the Pressure mode
does not affect the source setting made for the Flow mode and vice versa.
The nominal value (setpoint) for the Pressure mode can be defined internally (via keys) or
externally with a corresponding setpoint signal (0 … +10 VDC) fed via analog input AI 1.
The «Pressure» display shows the actual value
(pressure reading) and the nominal value (setpoint).
Press
to get to the
«Source Nom val Press» display.
The «Source Nom val Press» display shows the
setpoint source definition for the Pressure mode.
Press
to modify the source setting.
to save the new setting and return to the
«Pressure» display.
to return to the «Pressure» display.
SAVE
SAVE
Parameter Pressure
Act :
>1.00E+03mbar
Nom : 1.00E+00mbar
Min
: 1.33E+00mbar
Max : 1.33E+03mbar
SOURCE
PARAM SOURCE CONTROL
Pressure [ ]
Act :
mbar
Nom : 1.33E+01 mbar
>1.00E+03
SAVE
SAVE
Source Nom val Press
Source: internal