602494EA.DOCX
17200234_002_C1/2016 - © Leybold
27/102
Connector pin assingment X4 FCC68 receptacle
X4 sensor
Designation
Signal
Pin
Supply
+24 VDC
1
Supply common
GND
2
Signal input
0 … +10 VDC
3
Identification
4
Signal common
GND
5
Not connected
6
Not connected
7
Emission ON (HV ON)
14
+24 VDC
8
Table 4-4
Connector pin assingment DIN receptacle
X2 sensor
Designation
Signal
Pin
Identification
1
Supply common
GND
2
Signal input
0 … +10 VDC
3
Signal common
GND
4
Shielding
GND
5
Supply
+24 VDC
6
Table 4-5
14
Pressure sensors specified for a max. operating pressure (e.g. PTR 237) will start operation only after
the signal "Emission ON" is applied to the digital input DI 7 of the CMOVE 1250 (see chapter
«5.4 Operation Via Inputs and Outputs»).
1
8
1
2
3
4
5
6