GA10.218_1.01 - 11/2004
⏐
69
6.6.3 Mass
>
>
The requested mass of the measured gas can be selected. The Phoenix
must be in Stand-by for changing to another mass.
6.6.4 Interfaces
>
>
The parameters of the interface can be set.
6.6.4.1 Location of Control
>
>
>
Softkey 2: H
2
(2 amu)
Hydrogen with the mass of 2 amu will be measured.
Softkey 3:
3
He (3 amu)
Isotop of helium with the mass of 3 amu will be measured.
Softkey 7:
4
He (4 amu)
Helium with the mass of 4 amu will be measured.
Softkey 2:
Softkey 3: Define recorder output
Softkey 4:
Softkey 5: Define PLC outputs
Softkey 6: Define PLC inputs
Softkey 7:
Softkey 8: PLC sample rate
Softkey 2: PLC
The PhoeniX is controlled via the Digital In connector (See Chapter
). The START, STOP and ZERO buttons at the control panel and
remote control are locked.
Softkey 3: RS232
The PhoeniX is controlled via RS232 interface by an external computer. In
this mode the Phoenix can not be controlled via keyboard. The START,
STOP and ZERO button at the machine are deactivated.
Softkey 5: Local & PLC
The PhoeniX is controlled via the Digital In connector or/and the START,
STOP and ZERO buttons at the control panel and remote control.
Softkey 6: Local & RS232
The PhoeniX is controlled via the Digital In connector or/and the START,
STOP and ZERO buttons at the control paneland remote control.
Содержание PhoeniXL300
Страница 18: ...General Information 18 GA10 218_1 01 11 2004 ...
Страница 30: ...Operation Check 30 GA10 218_1 01 11 2004 ...
Страница 44: ...Working Principle 44 GA10 218_1 01 11 2004 ...
Страница 48: ...Operation 48 GA10 218_1 01 11 2004 refer to Chapter 6 2 1 Scale linear logarithmic Fig 15 Display Trend Mode ...
Страница 82: ...Menu 82 GA10 218_1 01 11 2004 ...
Страница 86: ...Calibration 86 GA10 218_1 01 11 2004 ...
Страница 98: ...Maintenance 98 GA10 218_1 01 11 2004 ...
Страница 101: ......