6.6.4.3 RS232 Protocoll
>
>
>
The calibration function of Leak Ware is not suitable with the use of a
Phoenix.
6.6.4.4 Define PLC outputs
The following relay outputs are available for further signal processing. The
maximum rating for the relay contacts is 60V AC/1A.
The contacts are numbered from bottom to top.
Description of the operation mode of the Digital Out. The pin assignement
for contacts 8 to 16 follows the same order as for pins 5 to 7.
The following digital out signals are selectable.
Trigger 1:
Is open in case Trigger Level 1 is exceeded or the machine is not in condi-
tion of measuring.
Trigger 2:
Is open in case Trigger Level 1 is exceeded or the machine is not in condi-
tion of measuring.
Trigger 3:
As Trigger 1 and Trigger 2
Softkey 2: Baudrate and end sign
Settings for the baudrate selectable between 1200 and 19200, as well as
endsign between CR+LF, CR or LF.
Softkey 3: Data, Parity, Stop bits
The settings for Data ( 7 or 8 ), Parity ( Even, Odd, None ) and Stop bits (
1 or 2 ) can be selected
Softkey 7: RS 232 Protocol
The protocol from the RS 232 can be choosen between: L200 size, Dia-
gnostic, L200 Leakware and ASCII code.
Pin
Assignment
1
PLC in free selectable
2
PLC in free selectable
3
PLC in free selectable
4
GND
5 to 7
Digital out free selectable, 5 center contact, 6 normally open
contact, 7 normally closed contact
8 to 10
Digital out free selectable
11 to 13
Digital out free selectable
14 to 16
Digital out free selectable
Содержание PhoeniXL300
Страница 18: ...General Information 18 GA10 218_1 01 11 2004 ...
Страница 30: ...Operation Check 30 GA10 218_1 01 11 2004 ...
Страница 44: ...Working Principle 44 GA10 218_1 01 11 2004 ...
Страница 48: ...Operation 48 GA10 218_1 01 11 2004 refer to Chapter 6 2 1 Scale linear logarithmic Fig 15 Display Trend Mode ...
Страница 82: ...Menu 82 GA10 218_1 01 11 2004 ...
Страница 86: ...Calibration 86 GA10 218_1 01 11 2004 ...
Страница 98: ...Maintenance 98 GA10 218_1 01 11 2004 ...
Страница 101: ......