2.5 Maximum Static Pressure for Pump Heads
Figure 9: Specification for maximum static pressure of pump head DCP-30
Note 1: Pressure limits only valid for pump head mounted on motor and do not account for tubing limits attached to the fittings.
Note 2: The above chart is also applicable to the DCP-30.2-SF1-G25 with sterile fittings.
2.6 Flow Control Specifications
Type
Flow Controller
Characteristics
PuraLev
®
iF30SU
with LFS-03SU-F1
PuraLev
®
iF30SU
with LFS-03SU-F1-SC1
PuraLev
®
iF30SU
with LFS-06SU-F1
PuraLev
®
iF30SU
with LFS-06SU-F1-SC1
Flow Range [l/min]
0
– 0.8
0
– 0.8
0
– 8
(7.7 for pump)
0
– 8
(7.7 for pump)
Accuracy of Reading
(at 20°C fluid temperature)
Note: Repeatability < Accuracy/2
> 35 ml/min: ±1%
< 35 ml/min:
> 6 ml/min: ±1%
< 6 ml/min: 0.06 ml/min
> 1.7 l/min: ±1%
< 1.7 l/min: ±17 ml/min
> 0.075 l/min: ±1%
< 0.075 l/min: 0.75 ml/min
Response Time:
Step from 10
– 90% of full scale.
< 1s
1
< 1s
1
< 1s
1
< 1s
1
Fluid Temperature / Ambient Temperature
Normal range: 10
– 60 °C ( 50 – 140 °F )
/
0
– 40
0
C (32
– 104
0
F)
Table 6: Specifications of flow controller systems
Note 1: Values for flow control parameters, which are optimized to the specific hydraulic circuit.
Default parameters are tuned for general flow control stability and may have higher values.
Figure 10: Accuracy and repeatability for single-use sensor LFS-03SU-F1
0
10
20
30
40
50
60
70
80
77
87
97
107
117
127
137
0
1
2
3
4
5
6
25
30
35
40
45
50
55
60
[psi]
[°F]
[bar]
Liquid Temperature in [°C]
Max. Static Pressure for DCP-30
Safety Factor = 1.3 (according to EN12162)
Note: Burst pressure is significantly higher.
Ambient Temp = 25
C
Note: Pressure limit only valid for pumphead mounted on motor. Values
only valid after Gamma radiation doses of < 40 kGy.
0%
1%
2%
3%
4%
5%
6%
7%
8%
9%
10%
1
10
100
1000
[ml/min]
Accuracy of Reading
Repeatability of Reading
35 ml/min Line
35 ml/min