User Manual for LEVIFLOW
®
LFS Flowmeters
www.levitronix.com
PL-4502-00, Rev11, DCO# 18-107
20
3.2.4 Instructions for Mechanical Installation
!
WARNING
Hazardous voltage may be present.
In order to avoiding fluid spills shorting voltages within the converter,
place it in a spill protected environment (for example protected
electronic cabinets).
If explosive flammable gases are present, place the converter in an
explosion-proof cabinet.
1. The converter shall be installed in a spill protected cabinet.
2. Assure that sufficient ventilation is provided in order to avoid exceeding the allowed ambient
temperature and humidity range:
➔
Temperature range: 0
– 50 °C (32 – 122 °F)
➔
Humidity range: 30 - 85% R.H. (no condensation)
3. Bring the fixation bars in to un-lock position (see
4. If multiple converters are used (see
) stack them together.
5. Attach the connector box (see
)
6. Attach the termination box (
) and fix it by pushing the fixation bars of the
Figure 17b and Figure 11
) to the inside.
7. Bring the fixation bars into the locked position (see
and
)
8.
Mount the whole stack on a DIN rail following the same steps as described in
the single channel converter
LFC-1C
.