3.1.4 Instructions for Mechanical Installation
!
WARNING
Hazardous voltage may be present.
In order to avoiding fluid spills shorting voltages within the converter,
place it in a spill protected environment (for example protected
electronic cabinets).
If explosive flammable gases are present, place the converter in an
explosion-proof cabinet.
1. The converter shall be installed in a spill protected cabinet.
2. Assure that sufficient ventilation is provided in order to avoid exceeding the allowed ambient
temperature and humidity range:
➔
Temperature range: 0
– 50 °C (32 – 122 °F)
➔
Humidity range:
30 - 85% R.H. (no condensation)
3. Mount the converter on a DIN rail according to the following steps:
Step 1:
Align first the
Snap-Pin
side of the converter
with the DIN rail corner (picture
a
). If the rigid
side is aligned first (see picture
b
) than some of
the DIN rail parts of the converter housing might
be damaged.
Step 2:
Press in the converter on the spring-pin side
(see picture
c
).
Step 3:
Press in the fix side of the DIN rail recess of the
converter housing (see picture
d
)
Step 4:
Relax the converter and let the snap-pin fix it
(see picture
e
).
Figure 16: Din Rail mounting steps for converter LFC-1C
(For dis-mounting follow the steps in reverse order)