19. GUIDANCE AND MANUFACTURER’S DECLARATION
19.1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is
used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations / Flicker
emissions IEC61000-3-3
Complies
Warning:
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed to
verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the pump, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
19.2 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is
used in such an environment.
Basic EMC standard
Immunity Test Levels
Compliance
Levels
Electromagnetic
Environment-Guidance
Professional
healthcare
facility environment
HOME HEALTHCARE
ENVIRONMENT
Electrostatic
Discharge (ESD)
IEC61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/ burst
IEC61000-4-4
±2kV for power supply line
±1kV for input/output line
±2kV for power
supply line
±1kV for input/
output line
Mains power quality should
be that of atypical commer-
cial or hospital environment
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to
line(s)
± 1 kV line(s) to
line(s)
Mains power quality should
be that of atypical commer-
cial or hospital environment.
Voltage dips, short
interruptions and vol-
tage variations on po-
wer supply input lines
IEC61000-4-11
Voltage Dips:
i) 100% reduction for 0.5 period,
ii) 100% reduction for 1 period,
iii) 30% reduction for 25/30 period,
Voltage Interruptions:
100% reduction for 250/300 period
230V
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of this device
requires continued operation
during power mains inter-
ruptions, it is recommended
that the device be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz) ma-
gnetic field
IEC61000-4-8
30 A/m
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.