![Levenhuk 74008 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/levenhuk/74008/74008_user-manual_1885319009.webp)
9
• In the eyepiece of the auxiliary microscope, besides the phase ring, the diaphragm ring can be seen.
• Center the image of the condenser annular diaphragm in respect to the objective phase ring using the adjustment screws
and then focus on the objective phase ring.
• Fully align the dark ring with the light one. If necessary, move the condenser upwards to align and center the rings with
each other and get the most contrast image. Use filters to make the image even more contrast.
• Replace the auxiliary microscope with an eyepiece and start researching.
• Note: Each change of a specimen or objective lens requires checking and/or centering of the annular diaphragm regarding
the phase ring.
Fuse replacement
Unplug the microscope from a power source. Open the fuse holder located on the back of the microscope body with a flat head
screwdriver. Remove the old fuse and install a new one. Only use the appropriate types of fuses. Close the fuse holder.
Digital camera
Levenhuk MED D45T comes with a digital camera (fig. 2a), Levenhuk MED D45T LCD comes with a digital camera with an LCD
screen (fig. 2b). A digital camera is installed to the third eyepiece tube and enables observing specimens on the screen in
exquisite detail and true color, taking photos and videos, and saving the images for future reference.
Specifications
misaligned rings
aligned rings
Product ID
74008
74009
74010
74011
Model
MED 45B
MED 45T
MED D45T
MED D45T LCD
Type
biological/optical
biological/optical, digital
Research method
phase contrast, bright and dark field
phase contrast, bright and dark field
Magnification
40—1000x
40—1000x
Interpupillary distance
48—75mm
48—75mm
Eyepiece head
binocular,
360° rotatable,
inclined at 30°
trinocular,
360° rotatable,
inclined at 30
trinocular, 360° rotatable,
inclined at 30°
Optics material
optical glass with an anti-fungal coating
optical glass with an anti-fungal coating
Eyepiece tube diameter
30mm
30mm
Third vertical eyepiece tube
—
23.2mm
23.2mm
Eyepieces
WF10x/22mm (2pcs)
WF10x/22mm (2pcs)
Eyepiece diopter adjustment
±5D
±5D
Objectives
Infinity Plan Phase contrast:
4x, 10x, 40xs, 100xs (oil)
Infinity Plan Phase contrast:
4x, 10x, 40xs, 100xs (oil)
Revolving nosepiece
5 objectives
5 objectives
Stage moving range
75/50mm
75/50mm
Stage
double layer mechanical, 180x150mm,
with a mechanical scale
double layer mechanical, 180x150mm,
with a mechanical scale
Focus system
coaxial, coarse (0.5mm)
and fine (0.002mm),
with rack and pinion mechanism
coaxial, coarse (0.5mm)
and fine (0.002mm),
with rack and pinion mechanism
Condenser
phase (dark field) with iris diaphragm,
with dovetail mount
phase (dark field) with iris diaphragm,
with dovetail mount
Body
metal
metal
Illumination
lower (5W LED) with brightness
adjustment
lower (5W LED) with brightness
adjustment
Collector
Köhler illumination
Köhler illumination
Filters
blue, green, yellow