Online commands
Leuze electronic
BCL 8
67
TN
T 3
5/7-
24V
Copy parameter set
Request parameter set from the BCL 8
Command
’PC’
Description
The command copies complete parameter sets.
Parameter
’03’ Copy parameters from the EEPROM into the RAM and initialize all
associated functions
'20' Copy standard parameters from the FLASH into the EEPROM and
RAM and initialize all relevant functions
’30’ Copy parameters from the RAM into the EEPROM
Acknowledg
-
ment
’PSx’
x
Status
'0' valid transmission
'1' invalid message
'2' invalid length of message
'3' invalid block check type
'4' invalid block check checksum
'5' invalid data length
'6' invalid message data
'7' invalid start address
'8' invalid parameter set
'9' invalid parameter set type
Example
’PC20’ loads the default parameters
Command
’PR’
Description
The command requests parameter data from the BCL 8. The <PS type>
parameter indicates from which parameter set the data are to be trans
-
ferred
Parameter
<BCC type> <PS type> <Start address> <Data length>
Acknowledg
-
ment
’PSx’
x
Status
'0' valid transmission
'1' invalid message
'2' invalid length of message
'3' invalid block check type
'4' invalid block check checksum
'5' invalid data length
'6' invalid message data
'7' invalid start address
'8' invalid parameter set
'9' invalid parameter set type
Example
’PR00102004’
Beginning with address 102, four (004) bytes are read out and transmitted