Programming functions
- 25 -
f) Default values programming
This function enables you to permanently program limits and settings so they are
maintained even after switching the instrument ON/OFF. This results in a
customised instrument.
The setting can only be changed in the respective functions (central switch position).
STORE user defined Parameters:
a) Continuity Level
Switch to Ohm position.
Press button.
for continuity and afterwards
.
Big display shows
.
The small display shows the actual continuity trigger level in a range from 0.1
to 50.0
Ω
. Incrementing steps are 0.1
Ω
within the range of 0.2
Ω
up to 5
Ω
-
from 5
Ω
to 50
Ω
stepping is 1
Ω
. Change the continuity level with the
key
and select the digit with
.
key.
Confirming this display with the
exits and saves.Turn rotary switch
without saving.
b) Limit value programming
Switch to desired measuring function.
Press the
key twice - large display shows
.
The small display shows the
LIMIT MIN
value for the selected function (rotary switch
position). Use
key to switch display to
. In this display
has no
function. Press
again to the show LIMIT
Change the LIMIT MIN with the buttons
(left most digit begins to flash) and
The entry is completed by pressing the
button or pressing
button to
store LIMIT MIN and exit.
Press the
key, the large display shows
the
key wraps arround
between HI and LO limit.The small display shows the
LIMIT MAX
value or OFF
(same as above).
Change the LIMIT MAX with the buttons
and
.
The entry is completed by pressing the
button. Values are stored permanently.
Содержание ISO plus
Страница 10: ...Getting started Operating principles 10 ...
Страница 11: ...Basic operation 11 5 Basic operation 5 1 Insulation test ...
Страница 13: ...Basic operation 13 5 2 Resistance measurement Continuity ...
Страница 15: ...Basic operation 15 5 3 Voltage measurement ...
Страница 17: ...Basic operation 17 5 4 Current measurement ...