Advanced functions
- 22 -
6.4 LIMIT
(Alarm)
The limit (alarm) function is activated with the
key and can
be used to detect when a value exceeds or falls below user
definable thresholds.
Exceeding the limit values is reported on the display and by
the buzzer (the lower limit with display "<LIMIT" and low
frequency sound, the upper limit with display ">LIMIT" and
high frequency sound).
The limits can be changed by pressing the
key again.
“Crossed limits” are supported for testing if a value is within a
certain range. The limit and appropriate >/< symbols are
displayed. They are not flashed. If the upper limit value is
placed below the lower limit value, the measured values
between the two limits are displayed with both annunciators
“<>" on and reported acoustically.
Limit values are displayed in the smaller display when in the
change limit mode. The limit values can be changed using
the
and
arrow keys. The high or low limit can be
turned off completely by pressing the the
key after
activating the limit display – the display shows OFF..
Pressing
swaps between the lower and upper limits.
Pressing
exits limit mode and saves any changes.
Message <LIMIT
Message >LIMIT
(low sound)
(high sound)
lower
limit value
upper
limit value
- full scale
value
+ full scale
value
Message ><LIMIT (sound)
upper
limit value
lower
limit value
- full scale
value
+ full scale
value
Содержание ISO plus
Страница 10: ...Getting started Operating principles 10 ...
Страница 11: ...Basic operation 11 5 Basic operation 5 1 Insulation test ...
Страница 13: ...Basic operation 13 5 2 Resistance measurement Continuity ...
Страница 15: ...Basic operation 15 5 3 Voltage measurement ...
Страница 17: ...Basic operation 17 5 4 Current measurement ...