32
Confocal Operation with Multi-photon
Detector Filters and Barrier Filter
The Super HyD detectors have a 620 nm barrier filter in front of them to protect against reflected IR
light. For bandwidth selection, you will need a suitable filter in front of the detector or detectors. Ask us
how to set this up.
Start Up
It is not necessary to power down the computer or the microscope when switching from
MP_LASER_OFF to MP_LASER_ON operation. However, certain steps must be followed for the
correct sequence.
Switch on the Electro-Optical Modulator.
The modulator is ready for use immediately, but does not
achieve full modulation depth for about 15 minutes. Switch it on
first
.
Switch off the Compact Supply Unit at the back of the table.
Switch on the Super-HyD detectors power and cooling unit.
Switch the Compact Supply Unit back on.
Application Start -
Select
MP_LASER_ON.