17
At right, the
Capture Image
button is used to acquire a single
image. The
Start
button is used to initiate acquisition of a stack
or a sequence.
Display
The displayed image is rotated 90° anti-clockwise with respect to the image seen in the objectives. Thus
x-axis stage translation moves the image vertically on the screen, and y-axis stage translation moves the
image horizontally.
Displaying Results
The look-up table LUT used to display the results may be
changed using the button at the top left of the display window.
The first click selects a range-finding LUT for the active window
that can be used to set gains and offsets. The second click
changes the display in all active windows. A third click returns
the LUT to the normal type.
The range-finding LUT is different to the one used by LSM. The LSM range-finding LUT used
blue for below range, red for above range, and white for within range.
below range
within range
above range
LAS AF uses green for below range, blue for above range, and red for within range.
below range
within range
above range