148
TC605/TC805/TC905/L-1.1.1en
© Leica
Remedy
If error occurs repeatedly,
contact service
department
Level the instrument or
re-adjust the V-index,
which adjusts also the
electronic level
Press
key
Bring target point within
tolerance range
Contact service
department
Error report
EDM System
TILT
Invalid Data
Out of Range
56
58
82
9x
Cause
System error in EDM
Total station not levelled
Invalid data for transfer to
TC605/TC805/TC905/L
For determination of
index- or collimation
error, V-angle is more
than ±6° (6.6666 gon) off
the horizontal plane
System defective
Remedy
Enter a new PtNr max. 8
digits
Change PtNr. manually
Change battery or connect
an external battery
Warning
PtNr Overflow
PtNr.Not Incr.
Battery Low
Cause
PtNr outside the upper
and lower limits
PtNr. cannot be
incremented
The battery is practically
flat (Level=0)
09
10
12