System Shutdown
1. Check the booking system to see if anyone is booked on the system within the
next hour. If not shutdown.
2. Select Configuration>Laser and drop the Argon slider back to standby
3. Close the LAS-AF program and shut down the PC
4. Turn the Laser Emission key to the horizontal position
5. DO NOT SWITCH OFF THE LASER POWER SWITCH UNTIL
THE ARGON FAN HAS STOPPED. Continue to next step while
waiting.
6. Switch off Scanner Power
7. Switch off PC/Microscope (Ensure PC is off before doing this)
8. Switch off the Mercury-Metal Halide Fluorescence Lamp via the power switch
Содержание SP5 confocal
Страница 7: ...7...
Страница 16: ...16 FRAP...
Страница 17: ...17...
Страница 18: ...18...
Страница 19: ...19...
Страница 20: ...20...
Страница 21: ...21...
Страница 22: ...22...
Страница 23: ...23...
Страница 24: ...24...
Страница 25: ...25 FRET Acceptor Photobleaching...
Страница 26: ...26...
Страница 27: ...27...
Страница 28: ...28...
Страница 29: ...29...
Страница 30: ...30...
Страница 31: ...31...
Страница 32: ...32 FRET Sensitised Emission...
Страница 33: ...33...
Страница 34: ...34...
Страница 35: ...35...
Страница 36: ...36...
Страница 37: ...37...
Страница 38: ...38...
Страница 39: ...39...
Страница 40: ...40...
Страница 41: ...41...
Страница 42: ...42 Spectral Imaging...
Страница 43: ...43...
Страница 44: ...44...
Страница 45: ...45...
Страница 46: ...46...
Страница 47: ...47...
Страница 48: ...48...
Страница 49: ...49...
Страница 50: ...50...
Страница 51: ...51...
Страница 52: ...52...
Страница 53: ...53...
Страница 54: ...54...
Страница 55: ...55...
Страница 56: ...56...
Страница 57: ...57...
Страница 58: ...58...
Страница 59: ...59...
Страница 60: ...60...
Страница 61: ...61...
Страница 62: ...62...
Страница 63: ...63...
Страница 64: ...64...
Страница 65: ...65...