Instruction manual V 2.3 – 02/2009
4.
Setup
The specimen clamps (standard or
universal cassette clamp) can also be
fastened directly to the fixture (with-
out dovetail clamping system).
4.5 Fastening the specimen clamp directly to the fixture
Fig. 9
17
16
14
10
13
18
12
12
•
Set specimen clamp (
12
) onto fixture for
specimen clamps (see
Fig. 9.3
) and fasten
with two screws (
17
).
•
Remove the two screws (
17
) (use Allen key
size 3) in order to detach the dovetail tenon
(
13
) from the specimen clamp (
12
) (
Fig. 9.2
).
•
Move fixture (
10
) to uppermost position
(turning the handwheel) and activate the
handwheel locking mechanism.
•
Remove dovetail clamping device (
14
) from
the fixture (
Fig. 9.1
).
To do so, loosen the four screws (
16
) with an
Allen key size 3 (
18
) and remove them.
To do so, proceed as follows: