Leica MC170 HD or Leica MC190 HD User Manual
Application 32
General notes
The white balance ensures that the
specimen is shown in neutral colors. We
recommend carrying out a new white balance
adjustment whenever you change the illumina-
tion type or color temperature. This is particu-
larly necessary if you are working with halogen
light that can be switched from yellow (low
intensity) to blue (high intensity).
Configure the white balance via the user
menu using the remote control. Depend-
ing on the configuration, automatic white
balance can also be carried out using the foot-
switch.
The white balance of the camera is
factory set to automatically provide opti-
mum results when used in conjunction with a
Leica LED illuminator.
Manual white balance
O
Manual white balance is described on
page 38.
Automatic white balance
1. Position a gray chart or a gray, neutral
object under the microscope so that the
entire image field is filled in.
2. Adjust the illumination as desired.
3. Press the button
on the remote control.
White Balance
Содержание MC170 HD
Страница 1: ...Leica MC170 HD Leica MC190 HD User Manual...
Страница 4: ...Leica MC170 HD or Leica MC190 HD User Manual General Notes 4 General Notes...
Страница 10: ...Leica MC170 HD or Leica MC190 HD User Manual Introduction 10 Introduction...
Страница 15: ...Leica MC170 HD or Leica MC190 HD User Manual Assembly 15 Assembly...
Страница 17: ...Leica MC170 HD or Leica MC190 HD User Manual Installation 17 Installation...
Страница 21: ...Leica MC170 HD or Leica MC190 HD User Manual Instrument Setup 21 Instrument Setup...
Страница 27: ...Leica MC170 HD or Leica MC190 HD User Manual Application 27 Application...
Страница 35: ...Leica MC170 HD or Leica MC190 HD User Manual User Menu 35 User Menu...
Страница 49: ...Leica MC170 HD or Leica MC190 HD User Manual Service 49 Service...
Страница 54: ...Leica MC170 HD or Leica MC190 HD User Manual Specifications 54 Specifications...