Service manual Leica M500 N, M520, M525
Service programs
Doc Code 10 665 751
3-1
08/2007 GAHE
Stock n
°
.
Description / Specification
Remark
Drawing
Service programs – Leica M500N, M520, M525
Service programs in control unit
How to enter the service functions:
•
Enter the service menu of the M500 N/M520 control unit in menue II
•
Enter security code: 6642
•
Select the service functions
Validity
Below description refers to SW 1.87 in the control unit and 2.4 or 4.4 in optics carrier.
Service function
Description Remark
Standard and user configuration
Settings of user- and standard
configuration can be changed and
stored.
Refer to
stand user manual
.
Ad. Function 1 output configuration The output of additional function 1 can
be set and stored either to toggle or to
switch mode.
Switch mode: relais contact closes
as long as the release button is
pressed.
Toggle mode: relais contact
opens/closes with each activation of
the release button.
Ad. Function 2 output configuration The output of additional function 2 can
be set and stored either to toggle or to
switch mode.
-“-
Screen saver setting
Screen saver for control unit display:
after the selected time the brightness
will be reduced.
Adjustable between 0..60 minutes.
Default = 5 minutes.
0 minutes will disable screen saver.
User parameter save time setting
When user settings are saved, the
display shows “User setting saved at…”
for a certain time; this time can be
altered.
Adjustable between 2..6 seconds.
Default = 4 seconds.
Power off timer setting
A power off / power on within the
selected time will effect:
The last selected user (before switching
off) is automatically taken as the actual
user.
Adjustable between 0..60 minutes.
Default = 5 minutes.
Содержание M500 N
Страница 2: ...Release 09 2007 First edition...
Страница 4: ...Description Service manual M500 N M520 M525 08 2007 Gahe 1 II Doc Code 10 665 751...
Страница 18: ...Description Service manual M500 N M520 M525 08 2007 Gahe 1 14 Doc Code 10 665 751...
Страница 20: ...Wiring Service manual M500 N M520 M525 07 2007 Gahe 2 II Doc Code 10 665 751...
Страница 23: ...Service manual M500 N M520 M525 Wiring Doc Code 10 665 751 2 3 07 2007 Gahe Iris switch assembly...
Страница 25: ...Service manual M500 N M520 M525 Wiring Doc Code 10 665 751 2 5 07 2007 Gahe LCD module Switch panel J12 P11...
Страница 29: ...Service manual M500 N M520 M525 Wiring Doc Code 10 665 751 2 9 07 2007 Gahe...
Страница 31: ...Service manual M500 N M520 M525 Wiring Doc Code 10 665 751 2 11 07 2007 Gahe Power supply removed since 03 2005...
Страница 40: ...Fault finding Service manual M500 N M520 M525 08 2007 Gahe 4 II Doc Code 10 665 751...
Страница 52: ...Fault finding Service manual M500 N M520 M525 08 2007 Gahe 4 14 Doc Code 10 665 751...
Страница 54: ...Replacing modules Service manual Leica M525 04 2004 3241 EBTH update 07 2007 5 II Doc Code 10 665 751...
Страница 64: ...Replacing modules Service manual Leica M525 04 2004 EBTH Update 07 2007 5 10 Doc Code 10 665 751...
Страница 66: ...Maintenance Service manual M500 N M520 M525 07 2007 Gahe 6 2 Doc Code 10 665 751...
Страница 68: ...Inspection Service manual M500 N M520 M525 07 2007 Gahe 7 II Doc Code 10 665 751...