background image

 

 

Leica M220 F12 / Ref. 10 716 922 / Version 03 1

Contents

1 Introduction 

2

1.1 

About this user manual/installation manual 

2

1.2 

Symbols in this user manual 

2

Product identification 

2

2.1 

Optional product features 

2

Safety notes 

3

3.1 

Intended use 

3

3.2 

Indication for Use 

3

3.3 Contraindications 

3

3.4 

Directions for the person responsible for the 
instrument 3

3.5 

Directions for the operator of the instrument 

4

3.6 

Dangers of use 

6

3.7 

Signs and labels 

8

4 Design 

10

5 Functions 

11

5.1 Illumination 

11

5.2 

Balancing system 

11

5.3 Footbrakes 

11

6 Controls 

12

6.1 

Horizontal arm 

12

6.2 

Tilt head/focusing unit 

12

6.3 

Footswitch (standard configuration) 

13

6.4 Stand 

13

6.5 

Optics carrier 

14

6.6 

Binocular tube, eyepiece, second observer tubes  14

Preparation before surgery 

15

7.1 Transportation 

15

7.2 

Positioning surgical microscope over operating 
table 16

7.3 

Installing the binocular tube, eyepiece and 
objective 17

7.4 

Installing the adapter for accessories 

18

7.5 

Adjusting second observer tube 

19

7.6 

Installing documentation accessories 

19

7.7 

Selecting documentation accessories 

20

7.8 

Adjusting the eyebase and eyepoint (optional)  21

7.9 

Adjusting the parfocality 

21

7.10  Changing accessories of the surgical microscope  

and balancing the swing arm 

22

7.11  Attaching the sterile components  

22

7.12  Starting up the surgical microscope 

22

7.13  Checks before operation 

23

8 Operation 

24

8.1 

Positioning the microscope 

24

8.2 

Adjusting the focus 

25

8.3 

Adjusting the magnification 

25

8.4 

Adjusting the illumination 

25

8.5 

Adjusting the working distance 

26

8.6 Decomissioning 

26

Components and accessories 

27

9.1 

Observer side 

27

9.2 

Video accessories for Leica M220 F12 

29

9.3 

Load table 

30

10 

Care and maintenance 

31

10.1  Care instructions 

31

10.2 Maintenance 

31

10.3  Care and Maintenance of the Leica Footswitch  31
10.4  Changing the fuse 

32

10.5  Replacing the LED module 

32

10.6  Checking the function of the LED illumination   32
10.7  Diagnostics in case of malfunctions 

33

10.8  Notes on reprocessing of resterilizable products  34

11 Disposal 

36

12 

What to do if ...? 

36

12.1  General malfunctions 

36

12.2 Microscope 

36

12.3  TV, photography 

37

13 

Technical data 

38

13.1  Electrical data 

38

13.2  Surgical microscope 

38

13.3 Stand 

38

13.4  Optical data 

39

13.5  Control unit  

39

13.6  Configurations and Weights 

40

13.7  Ambient conditions 

40

13.8  Electromagnetic compatibility (EMC) 

40

13.9  Standards fulfilled 

41

13.10  Limitations of use 

41

13.11 Dimensions 

42

Содержание M220 F12

Страница 1: ...MEDICAL DIVISION User Manual Installation Manual 10 716 922 Version 03 Leica M220 F12...

Страница 2: ...rugg Tel 41 71 726 3333 Fax 41 71 726 3334 Legal disclaimer All specifications are subject to change without notice The information provided by this manual is directly related to the operation of the...

Страница 3: ...t optional 21 7 9 Adjusting the parfocality 21 7 10 Changing accessories of the surgical microscope and balancing the swing arm 22 7 11 Attaching the sterile components 22 7 12 Starting up the surgica...

Страница 4: ...not avoided may result in minor or moderate injury Note Indicates a potentially hazardous situation or improper use which if not avoided may result in appreciable material financial and environmental...

Страница 5: ...t regular inspections to make certain that the authorized users are adhering to safety requirements X When instructing new users do so thoroughly and explain the meanings of the warning signs and mess...

Страница 6: ...used no closer than 30 cm 12 inches to any part of the Leica M220 including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Note The EMISS...

Страница 7: ...ed on has within the last 24 hours already been exposed to illumination from the same or any other ophthalmological instrument that uses a bright visible light source This applies especially to patien...

Страница 8: ...all preparations and adjustments on the optics carrier before the operation X Never rebalance or re equip the optical components and accessories with the instrument over the operation area X Before re...

Страница 9: ...neath the base of the stand X Always push the surgical microscope never pull it CAUTION Surgical microscope can move without warning X Always secure the footbrakes when you are not moving the system C...

Страница 10: ...Safety notes 8 Leica M220 F12 Ref 10 716 922 Version 03 3 7 Signs and labels 1 11 12 10 2 6 3 9 8 4 5 6 7 15 10 13 14 10 10 10 10...

Страница 11: ...tch e JJ year 2 digits f MM month 2 digits g TT day 2 digits h Production starting date 4 Type label 5 Follow the user manual 6 Optical radiation 7 Symbol for fuses 8 Transport position F12 floor stan...

Страница 12: ...on 03 4 Design 1 2 5 4 6 7 3 8 14 14 9 12 13 10 11 1 Swing arm 2 XY unit optional 3 Focus unit 4 Tilt head 5 Eyepieces 6 Binocular tube 7 Optics carrier 8 Objective 9 Horizontal arm 10 Column 11 Foots...

Страница 13: ...without tilting or falling down After balancing all movements during operation only need a minor force With the rotary knob 3 on the swing arm 2 the drifts will be adjusted The brake knob 1 locks the...

Страница 14: ...tch Only connect the footswitch provided by Leica Microsystems Schweiz AG Medical Division 2 Power switch 3 Power connection 4 Cable holder for footswitch cable 5 Power cable strain relief device 6 2...

Страница 15: ...ication change The assignment of 1 and 3 magnification and focus control can be switched For additional information refer to the service manual or ask your local after sales service 6 4 Stand 1 3 2 1...

Страница 16: ...ection filter GG475 GG435 and protection filter 5 5 5 Emergency drive for the magnification changer In case of emergency the magnification can be adjusted on the emergency drive using a coin or the li...

Страница 17: ...microscope can move without warning X Always secure the footbrakes when you are not moving the system CAUTION Surgical microscope can tilting without warning X Always push the surgical microscope in...

Страница 18: ...zontal arm and guide it through the cable guide 3 1 2 3 CAUTION Danger of infection X Leave sufficient space around the stand to prevent contact with non sterile components X Inspect all terminals and...

Страница 19: ...ely Inclined binocular tube Inclined binocular tube 45 Optional for use on the assistant attachment not a standard configuration Binocular tube 5 25 Optional for use on the assistant attachment not a...

Страница 20: ...20 F12 features two slots for additional filters You can use temperature resistant filters with a diameter of 32 mm For further information please contact your Leica Microsystems representative 1 X Pu...

Страница 21: ...tting at the eyepiece X Correct the image with the knurled ring 2 7 6 Installing documentation accessories WARNING Risk of injury from parts falling down X Complete all preparations and adjustments on...

Страница 22: ...TV dual attachment Zoom video adapter TV attachment 70 mm 1 2 3 1 2 1 3 1 4 107 mm 100 mm 85 mm 60 mm 55 mm 35 mm Photo TV dual attachment TV attachment TV attachment TV attachment Photo TV dual attac...

Страница 23: ...e separately 1 only this method will ensure that the image will stay in focus throughout the entire magnification range parfocal Prepare the microscope X Switch on the microscope at the control unit a...

Страница 24: ...the swing arm X Hold the microscope firmly X Loosen the brake knob for locking the vertical position 2 The swing arm is now released X By moving the swing arm manually determine whether more force is...

Страница 25: ...le replacement LED module ready 1 X Switch on the microscope using the power switch The LED is illuminated X Turn the button 1 to the right This increases the brightness X Test the illumination across...

Страница 26: ...eached the LED 1 lights up X Using the footswitch move the focus into reset position see mark 3 Coarse positioning X Hold the microscope by the two handles 4 and position it Fine positioning X Actuate...

Страница 27: ...X On the lower potentiometer 3 of the board set the speed to the desired value 3 8 3 Adjusting the magnification X Set the magnification on the footswitch using button 1 chapter 8 2 The following valu...

Страница 28: ...scope using the handles 1 X Set the exact working distance using the button on the footswitch for the focus drive 8 6 Decomissioning X End the recording procedure on the camera loss of data X Return t...

Страница 29: ...ical microscope swinging down X Do not exceed the max load when equipping components and accessories X Check the total weight using the Load table chapter 9 3 9 1 Observer side 1 2 2 4 Main surgeon As...

Страница 30: ...3 Binocular tube inclinable 5 to 25 with PD Adjustable viewing angle and height Adjustable interpupillary distance 3 Binocular tube inclined T type II 3 Binocular tube inclined TC 3 Binocular tube 45...

Страница 31: ...eica HD C100 Video adapter For commercially available video cameras with C mount complete with adapter The video adapter 2 is installed at the beam splitter Zoom and fine focus function for Leica Zoom...

Страница 32: ...457297 Objective f 225 mm 10448217 Binocular tube inclinable 5 to 25 with PD 0 74 kg 10446574 Binocular tube inclinable T type II 0 74 kg 10446618 Binocular tube 45 0 56 kg 10448404 Binocular tube inc...

Страница 33: ...litting compounds strong organic acids oxygen splitting compounds X Follow the disinfectant manufacturer s instructions It is recommended to conclude a service contract with Leica Service Tropical env...

Страница 34: ...all the way using your hand 10 5 Replacing the LED module CAUTION Burn hazard The LED module get very hot X Check that the LED module have cooled before you remove it The Leica M220 F12 does not have...

Страница 35: ...the footswitch X Press the button on the footswitch The corresponding LED 1 on the board in the FOOT SWITCH FUNCTIONS area must be illuminated If the LED is not illuminated the footswitch may be defec...

Страница 36: ...Alcohol may also be used to clean the optics if heavy contamination such as fingerprints grease streaks etc is present X Dry off products except for optical components with a disposable cloth paper cl...

Страница 37: ...Clip on handle 10428328 Rotary knob binocular tubes T 10384656 Rotary knob transparent 10443792 Lever extension 10446058 Protective glass multifocal lense 10448439 Protective glass 10448440 Cover ster...

Страница 38: ...effort Articulation brakes set too tightly X Loosen the articulation brakes see chapter 7 1 The functions cannot be actuated via the footswitch The cable connection has loosened X Check the power cabl...

Страница 39: ...out diagnostics according to chapter 10 7 In the event of a defect contact Leica Microsystems service 12 3 TV photography Malfunction Possible cause Remedy The image on the monitor is too dark The vid...

Страница 40: ...ed reflex illumination Light source Direct and long lasting LED illumination Average service life of 60 000 h for an end of life criterion of 70 of the initial brightness UV filter UV and IR free LED...

Страница 41: ...2 5 17 4 6 29 2 46 0 7 3 Objective f 200 mm Eyepiece Total magnification mm Field of view mm min max min max 8 33 22 2 7 17 0 68 1 10 8 10 21 3 2 20 4 65 0 10 3 12 5 17 4 0 25 5 52 6 8 3 Objective f 2...

Страница 42: ...t for near active HF Surgical Equipment and the RF shielded room of an ME System for magnetic resonance imaging where the intensity of EM Disturbances is high Compliance IEC 60601 1 2 Emissions CISPR...

Страница 43: ...Schweiz AG holds the management system certificates for the international standard ISO 13485 relating to quality management and quality assurance 13 10 Limitations of use The Leica M220 F12 surgical...

Страница 44: ...Technical data 42 Leica M220 F12 Ref 10 716 922 Version 03 13 11 Dimensions Unit of measure mm...

Страница 45: ...MEDICAL DIVISION Installation Manual Leica M220 F12...

Страница 46: ...1 Installing the F12 floor stand 46 2 2 Mounting the optics carrier 47 2 3 Installing the XY unit optional and optics carrier 48 2 4 Cabling 49 3 Checklist 50 4 Disassembly 50 5 Instrument Setup 51 5...

Страница 47: ...d nuts Tools NOTE Screws or threads can be destroyed X Tighten all screws without using excessive force Allen key 2 5 mm 5 mm Screwdrivers for Phillips head screws Size 0 Assembly material Cable ties...

Страница 48: ...the column into the base X Fasten it using the 2 pretreated Allen screws 1 provided X Close off the threaded holes 1 using the plastic plugs provided 2 X Attach the swing arm to the column X Check the...

Страница 49: ...rrows 2 on the swing arm and optics carrier must be positioned one above the other X Unscrew the 4 Phillips head screws 3 of the focus cover X Remove the focus cover 4 X Feed the cable from the swing...

Страница 50: ...d XY unit must match X Fasten the XY unit to the swing arm using the 4 pretreated Allen screws 10 provided The red arrows on the XY unit and the swing arm must be positioned one above the other 2 3 2...

Страница 51: ...asten the power cable 3 to the swing arm using cable ties 5 X Connect the footswitch cable 2 and guide it through the cable guide 4 2 4 2 Routing the video cable optional X Guide the video cable throu...

Страница 52: ...ca M220 F12 Ref 10 716 922 Version 03 3 Checklist Are all the cables routed correctly and not pinched 4 Disassembly X Disassemble the Leica M220 F12 surgical microscope in the opposite order in which...

Страница 53: ...eica M220 F12 and move the stand over the entire range of movement Result The brakes disengage the microscope can be moved over the entire range of movement freely and without noise X Move down the X...

Страница 54: ...d acceptance program This inspection is aimed at determining whether the safety requirements concerning the protection of the patients and of the medical personnel have been fulfilled the performance...

Страница 55: ...ntained in the assembly instructions Name and address of the customer Department Representative of Leica Microsystems Schweiz AG Medical Division Name Address Acknowledged Signature Location Date This...

Страница 56: ...trasse 201 CH 9435 Heerbrugg T 41 71 726 3333 F 41 71 726 3399 www leica microsystems com 10 716 922en 03Copyright byLeicaMicrosystemsGmbH 2021 Switzerland Subjecttomodifications Printed 01 2021LEICAa...

Отзывы: