Technical data
40
Leica M220 F12 / Ref. 10 716 922 / Version 03
13.6 Configurations and Weights
From the interface of the microscope, the stand has a maximum
load of 5.0 kg.
Determine the total weight of the load using the "Load
table",chapter 9.3.
13.7 Ambient conditions
Use
+10 °C to +40 °C
+50 °F to +104 °F
30 % to 75 % relative humidity
780 mbar to 1013 mbar air pressure
Storage:
–30 °C to +70 °C
–22 °F to +158 °F
10 % to 100 % relative humidity
500 mbar to 1060 mbar air pressure
Transport:
–30 °C to +70 °C
–22 °F to +158 °F
10 % to 100 % relative humidity
500 mbar to 1060 mbar air pressure
13.8 Electromagnetic compatibility
(EMC)
Environment for which the instrument is suitable
Hospitals except for near active HF Surgical Equipment and the RF
shielded room of an ME System for magnetic resonance imaging,
where the intensity of EM Disturbances is high.
Compliance IEC 60601-1-2
Emissions • CISPR 11, Class A, Group 1
• Harmonic Distortion per IEC 61000-3-2 Class A
• Voltage Fluctuation and Flicker per IEC 61000-3-3
Class A, Figures 3-7
Immunity • Electrostatic discharge IEC 61000-4-2:
CD +/- 8 kV, AD +/- 15 kV
• Radiated RF EM Fields IEC 61000-4-3:
80 – 2700 MHz: 10 V/m
• Proximity Wireless fields IEC 61000-4-3:
380 – 5785 MHz: 9 V/m; 28 V/m
• Electrical Fast Transients and
bursts IEC 61000-4-4:
± 2 kV: Power supply lines
• Surges IEC 61000-4-5:
± 1 kV Line-to-line
± 2 kV Line-to-ground
• Conducted disturbances, induced by RF fields
IEC 61000-4-6:
10 V rms
• Rated Power-frequency Magnetic Field IEC 61000-
4-8:
30 A/m
• Voltage dips and interruptions IEC 61000-4-11:
according to IEC 60601-1-2:2014